OUBLESHOOTIN

G

TROUBLESHOOTING

PROBLEM

PROBLEM

SOLUTION

SOLUTION

Tubing Out *

1.  Tubing sensor indicator is not seated

2.  Ensure that the tubing sensor coupler  

    is seated completely

3.  Change the tubing

*    Return to Arthrex

®

 for repair

1.  Ensure that no tubing is connected to the  

    pump during power on sequence

*   Return to Arthrex for repair

1.  Check the AC main power cord

2.  Try an alternate power outlet

3.  Check the AC main fuses

*   Return to Arthrex for repair

1.  Activate LAVAGE mode

2.  Increase pressure

3.  Decrease outflow

1.  Check for error messages

2.  Open all tubing clamps and shut-off

    valves

3.  Ensure actual pressure is below the

     target pressure

4.  Check if the tubing is pinched, kinked,  

    or blocked

*    Return to Arthrex for repair
1.  Roller housing not secured – ensure  

    locking lever is properly secured

*    Return to Arthrex for repair

1.  Increase pressure

2.  Activate LAVAGE mode

3.  Reduce outflow

4.  Use high-flow cannulas

Console fails 

Self Diagnostic 

Test*

Console will  

not power up

Lack of  

distention liquid 

bloody/cloudy

Does not pump 

when activated

Door Open*

Inadequate 

pressure

No/Low Flow

1.  Check for error messages

2.  Check that all tubing clamps are open

3.  Check the settings for flow and pressure

4.  Check if the tubing is pinched, kinked  

    or blocked

5.  Check that tubing seats correctly over  

    the rollers

6.  Verify use of high-flow cannulas

7.  If failure remains, return to Arthrex  

    for repair

1.  If failure remains, increase or open outflow

2.  Manipulate joint to stress-free position

1.  Ensure adequate fluid supply

2.  Decrease outflow

3.  Check tubing for damage and if it is    

    pinched, kinked or blocked

4.  Check tubing for proper connections

5.  Replace tubing

6.  If failure remains, return to Arthrex  

    for repair

Synergy Shaver Detection

1.  Verify Synergy Resection Console is  

    at least 2.3

2.  Ensure the Synergy cable is connected

3.  Press stop and then run on CW4 while  

    the Synergy Resection Console is on

4.  Replace the shaver detection cable

5.  If failure remains, return to Arthrex  

    for repair

Overpressure

Pressure Fault

No Shaver  

Detection

For a full listing of warnings and instructions, please consult the Directions for Use and User Manual, packaged with the pump or 
pump tubing. These materials can also be found on Arthrex.com.

Summary of Contents for CONTINUOUS WAVE

Page 1: ... 1945 USA Toll Free 1 800 934 4404 www arthrex com Arthrex GmbH Erwin Hielscher Strasse 9 81249 München Germany Tel 49 89 909005 0 www arthrex de Arthrex Tech Support 1 888 420 9393 49 89 90 90 05 8800 For more information consult the User s Guide and the Directions for Use supplied with the pump or tubing sets ...

Page 2: ...e of flow for improved control of debris removal Only available when using DualWave outflow Shaver Suction Toggle Allows switching be tween four preset shaver suction levels Only available when using DualWave outflow Lavage Pressure Decrease Shaver Suction Toggle Rinse Pressure Increase Run Stop Autoclavable Remote Foot Control ...

Page 3: ...sert the green tubing hub next to the green mark Wrap the thicker tubing between the hubs and around the rollers 1 6 8 2 7 3 4 5 Rollers Inflow door Tubing sensor coupler Clear tubing hub Close inflow latch Remote foot switch connector Power button a b d c Turn on the system and wait until the self diagnostic testing has finished before attaching the tubing Stretch the thicker tubing to insert the...

Page 4: ...W ONLY USER INTERFACE Stopped Running Menu Boost Control Set Pressure Pressure Decrease Lavage Pressure Increase Run Stop Run Stop Pressure Reading Run Time Resets after power down Inflow Only Fluid Used Flow Rate ...

Page 5: ...ergy imaging console for the heads up display Power on the Synergy Resection Console 2 1 4 Power on the CW4 3 For the Synergy heads up display use the AR 3200 1040 accessory cable kit to connect the CW4 to camera console Only 1 cable is needed to connect the CW4 to the camera console See IFU 950 0052 00 for more detail 2 3 NOTE Start with all units turned off CW4 Synergy ...

Page 6: ...NOTES ...

Page 7: ...oudy Does not pump when activated Door Open Inadequate pressure No Low Flow 1 Check for error messages 2 Check that all tubing clamps are open 3 Check the settings for flow and pressure 4 Check if the tubing is pinched kinked or blocked 5 Check that tubing seats correctly over the rollers 6 Verify use of high flow cannulas 7 If failure remains return to Arthrex for repair 1 If failure remains incr...

Page 8: ...judgment in making any final determinations in product usage and technique In doing so the medical professional should rely on their own training and experience and should conduct a thorough review of pertinent medical literature and the product s Directions For Use http www arthrex com imaging resection fluid management View U S patent information at www arthrex com corporate virtual patent marki...

Reviews: