950-0073-00 B SM, English, Arthrex Synergy UHD4 System
53 of 74
STERIS® and STERRAD® Protocols Validated
Sterilization
Protocol
Details
STERIS® Systems
AR-3210-XXXX Camera Heads are validated for Sterilization Assurance using the STERIS® systems
listed below, and following the manufacturer’s instructions.
V-PRO® 1 Low Temperature Sterilization System [Standard Cycle]
V-PRO® 1 Plus Low Temperature Sterilization System [Lumen and Non-lumen Cycles]
V-PRO® maX Low Temperature Sterilization System [Lumen, Non-lumen, and Flexible Cycles]
V-PRO® 60 Low Temperature Sterilization System [Lumen and Non-lumen Cycles]
STERRAD® Systems
AR-3210-XXXX Camera Heads are validated for Sterilization Assurance using the STERRAD® Systems
listed below, and following the manufacturer’s instructions for use.
STERRAD® System 100S [Short Cycle]
STERRAD® System NX [Short Cycle]
STERRAD® System 100NX [Standard Cycle]
4.3.5
Material Compatibilities
In addition to the Sterilization chemicals listed above, the AR-3210-XXXX camera heads are Material
Compatible with Cidex OPA. No SAL claims are made with Cidex OPA.
WARNING: Use of Sterilants or Chemicals other than those listed in the Cleaning and
Sterilization section may result in the compromise of the device’s safety and effectiveness. Use
of Sterilants or Chemicals other than those listed in the “Cleaning and Sterilization” section shall
void the product’s warranty.