Routine Care
6-10
Angel
®
cPRP System
Operator’s Manual
Fuse Replacement
In the event that a fuse fails, follow these steps to replace it. If the replacement fuse fails, there may
be a problem with the electrical outlet or an electrical component failure internal to the machine.
Contact Arthrex, Inc. Customer Service.
WARNING
To prevent risk of electrical shock, shut OFF power and unplug the system from the
electrical outlet before performing cleaning procedures or replacing the fuses.
1. Power Switch cover conceals the two fuses
Figure 6-10 Fuse Replacement: Pry Open Cover
1. Using a flat head screwdriver, pry open the cover on the Power Switch.
2. Open the cover. The two fuse holders are visible.
Figure 6-11 Fuse Replacement: Slide Out Holders
Note the orientation of the downward arrows on the fuse holder. Pull each holder to remove them
from the machine.