Programmability Option
Angel
®
cPRP System
Operator’s Manual
4-7
1. Protocol Buttons
Copy Button
Touching the Copy Button will duplicate the current protocol. This is the
only way to create a new protocol. The newly created protocol will have
a number prefix added to its name. Touch the Rename Button to rename the
new protocol.
Rename Button
Touch the Rename Button to rename the copy of a protocol. Touching the
Rename Button will display a keyboard on the Touch Screen Display.
Please refer to the illustration in Figure 4-2. Type in the new protocol name, and then
touch “
Enter
” to save and return to the “
Protocols
” tab.
Delete Button
Touching the Delete Button deletes the current protocol. Once deleted,
it will no longer be available from the Protocol Pull-Down Button.
Undo/Redo Buttons
Touching the Undo Button will undo your last action (e.g.,
changing the value of a protocol parameter or renaming a
protocol). Touching the Redo Button will redo the last action.
You may undo and redo up to ten different actions.
2. Protocol Parameters and Buttons
The protocol parameters consist of the six value fields.
Touching a parameter will highlight both the parameter and the up and down arrow
buttons. Use the arrow buttons to change the value of the selected parameter.
Every value field has a minimum and maximum value. The values are unable to be
changed beyond the minimum and maximum values.
Default volume to process (mL)
This protocol parameter sets the default volume of whole blood or the mixture of blood
and bone marrow to process in a single cycle.
Hct of PRP product (%)
This protocol parameter sets the amount (%) of Hematocrit contained in the PRP
product of a case. This value is based on a harvest of PRP (including PPP dilution)
equivalent to 10% of the whole blood or the mixture of blood and bone marrow
volume processed.
Summary of Contents for Angel
Page 2: ...This page intentionally left blank ...
Page 8: ...Overview vi Angel cPRP System Operator s Manual This page intentionally left blank ...
Page 14: ...Overview xii Angel cPRP System Operator s Manual ...
Page 17: ...Overview Angel cPRP System Operator s Manual xv This page intentionally left blank ...
Page 18: ......
Page 26: ...Overview 1 8 Angel cPRP System Operator s Manual This page intentionally left blank ...
Page 66: ...Processing 3 28 Angel cPRP System Operator s Manual This page intentionally left blank ...
Page 98: ...Routine Care 6 12 Angel cPRP System Operator s Manual This page intentionally left blank ...