Page 10 · 4200 Series
LaserSource
User’s Manual
Operation
The Front Panel
Operation of the
LaserSource
is very straightforward. The sections below will
help familiarize you with the front panel, the display, and the menu structure.
The front panel is designed for simplicity in operation. There are three buttons
on the front panel:
Output
,
Menu
, and the adjustment knob.
The output is controlled via the
Output
button. When turning the output on,
there is a programmable delay (factory default of 3 seconds) where the unit will
indicate that the output is on (the blue output LED will be lit), but the output
stage is in fact not on. This is a CDRH safety feature required for some laser
applications. The output will be energized once the delay has expired.
The
Menu
button is used to enter the
LaserSource
menu. When in remote
mode, it acts as a
Local
button, returning the instrument to local mode.
The large adjustment knob located on the right hand side of the unit is used to
change the set point or parameters in the menu. It also acts as a push button,
primarily as an enter button, when making changes in the menu.
There are four LEDs: a green power LED that is lit whenever the AC power is
turned on; the blue
Output
on LED that indicates the
LaserSource
is actively
controlling the laser diode; a red
Error
LED that indicates an error has occurred;
and a yellow
Remote
LED that indicates the unit is being controlled via a
computer.
Whenever an error is generated, the red Error LED will light, and the error will be
displayed on the VFD display. There may be one or more errors, but only the
first error will be displayed. To display the next error, press
Menu
button. To
clear all errors, press the knob. A list of error codes can be found in the
Error
Messages
section below.