66
Designs, materials, weights and performance ratings are approximate and subject to change without notice. Visit
armstrong
international.com for up-to-date information.
Operating Instructions – continued
Diagnostics Menu
Use the Diagnostics Menu to simulate operation and review the system files. The system log files contain time/date
stamped messages including: power on, power off, programming time outs, parameter faults, incorrect password entry
and other various information relative to system operation and programming.
The simulated inputs are for testing the meter to verify that the programming is correct. They are also used to enter
nominal operating temperature and pressure for the V1 only model. Simulated vortex frequency allows you to enter any
value for the sensor input in Hz. The meter will calculate a flow rate based on the corresponding value and update all
analog outputs
(the totalizer display and output is not affected by a simulated frequency)
. The simulated pressure
and temperature settings work the same way. The meter will output these new values and will use them to calculate a
new density for mass flow measurement. Note: when your diagnostic work is complete, make sure to return the values
to zero to allow the electronics to use the actual transducer values. For the V1 only model keep the temperature and
pressure at nominal operating conditions.
If the meter display indicates a temperature or pressure fault, a substitute value can be entered to allow flow calculations
to continue at a fixed value until the source of the fault is identified and corrected. The units of measure of the displayed
values are the same as the units configured for the flow meter.