Data Hold

 

Data Hold is not available when “OL K

” 

displayed with no input signals. 

 

Data Hold Beep Guard 

The internal sounder will operate  
continuously with LCD display flashing in 
two situations in the Data Hold mode: 
1. The Meter measure a signal different  
    from the LCD reading. 
2. The measured signal is the same unit  
    as the LCD reading and is larger 50  
    counts than the LCD reading. 

 

Auto Hold  
(ACA and DCA only)

 

Battery Replacement

 

Specifications 

1-1  General Specifications 

 

LCD display digits :  

9999 digit large scale LCD readout. 

Measuring rate : 

5 times / sec. 

Overrange display :  

“OL” is displayed for “

” and “V” function, 

shows the real value for “A” function. 

Automatic power off time :  

Approximately 20 minutes after power on 
without operations.  

Low battery indicator : 

<

 is displayed. 

Replace the battery when the indicator  

<

 appears in the display.

 

Power requirement : 

9V battery.  

Battery type and life :  

ALKALINE PP3, 9V 125 hours. 
 

1-2  Environmental Conditions 

 

Indoor Use. 

Calibration :  

One year calibration cycle. 

Operating temperature :  

0°C ~ 30°C (

80% RH)  

30°C 

40°C (

75% RH)  

40°C 

50°C (

45%RH) 

Storage temperature : 

-20 to +60°C,  

0 to 80% RH (batteries not fitted). 

Temperature coefficient :  

0.2 x (Specified accuracy) 

°C, < 18°C,  

> 28°C 

.

 

Operating altitude : 

2000m (6562 ft) 

Overvoltage category :  

IEC 61010-1 1000V CAT.

,  

                       600V CAT.

.  

CAT         Application field

 

1-3  Electrical Specifications 

Accuracy is 

±

(% r number of 

digits) at 23°C  

±

 5°C < 80%RH.  

(23°C  

±

 3°C < 80%RH  for ACA & DCA)  

 

Voltage

 

Function 

Range 

Accuracy 

V



 

1.3V ~ 750.0V  

±(

0.9%+ 3 dgt) 

50Hz ~ 60Hz 

±(

1.5%+ 3 dgt) 

61Hz ~ 500Hz 

2.1V ~ 1000V 

±(

0.3% + 2 dgt) 

-0.7V ~ -1000V 

V



 

Overload protection :  

AC 750Vrms // DC1000V 

Max Operation time :  

DT=30s for 

30V 

Input impedance :  

4K for input voltage up to 30V.  

Impedance increases with input voltage  
to approximate 375K

 

at 750V. 

AC Conversion Type : 

AC Conversions are ac-acoupled, true 
RMS responding, calibrated to the RMS 
value of a sine wave input. Accuracies 
are given for sine wave at full scale and 
non-sine wave below half scale. For non
-sine wave add the following Crest  
Factor corrections : 
For Crest Factor of 1.4 to 2.0, add 1.0% 
to accuracy. 
For Crest Factor of 2.0 to 2.5, add 2.5% 
to accuracy. 
For Crest Factor of 2.5 to 3.0, add 4.0% 
to accuracy. 
CF 3 @ 400A, 400V 
      2 @ 600A, 600V

 

 Caution  

The calibration seal can only be  
punctured by certified distributor for  
annual calibration otherwise the meter’s 
performance can not be warranted.

 

Conductor Size : 

35mm diameter.

 

Pollution degree : 

2  

EMC : 

EN 61326-1

   

 

 

The circuits not connected to mains. 

 

The circuits directly connected to  

Low-voltage installation. 

 

The building installation. 

 

The source of the Low-voltage 

installation. 

Shock vibration : 

Sinusoidal vibration 

per MIL-T-28800E (5 ~ 55 Hz, 3g  
maximum).

 

Auto Backlight

 