The following symbols are used in this operating manual to call your 
attention to specific items.

■ For your safety

Follow the instructions given here to prevent exposure to 
pathogenic microorganisms.

Follow the instructions given here to prevent injury and 
property damage.

■ For optimal performance

IMPORTANT:

Follow the instructions given here to obtain accurate measurement 
results.

NOTE:

Information useful for preventing damage to the instrument or parts, 
and other important information you should keep in mind.

REFERENCE:

Additional explanations that help you make the best use of the instrument 
and information on related functions.

This operating manual contains important information on the functions of 
The Lab 004.

This operating manual is issued by ARKRAY, Inc.
Read carefully prior to starting up the unit.
It is recommended to retain this operating manual for future use.

This product conforms to the EMC Standard IEC61326-2-6:2012 
(EN61326-2-6:2013).
Class of emission: CISPR 11 Class A
This instrument is an IVD medical instrument.

This product conforms to European Directive 98/79/EC.

NOTE: This instrument generates, uses and can radiate radio frequency 
energy and, if not installed and used in accordance with the operating 
manual, may cause harmful interference to radio communications. 
Operation of this instrument in a residential area is likely to cause 
harmful interference in which case the user will be required to correct 
the interference at his own expense.

The electromagnetic environment should be evaluated prior to operation 
of the device. Do not use this device in close proximity to sources of 
strong electromagnetic radiation, as these may interfere with proper 
operation.

Premise

Symbols

Caution Labels

Thank you for purchasing our compact biochemical analyzer, The Lab 
004.
Read this operating manual thoroughly before using the instrument. This 
operating manual gives an overview of the instrument and the proper 
procedures for operation and maintenance. Follow the instructions in 
this operating manual in order not to defeat the purpose of the protective 
features of the instrument.
For the purchase of reagents, consumables or other optional items, 
contact your distributor.

• TAKE THE UTMOST CARE WHEN HANDLING BLOOD. This 

instrument uses blood as sample. Blood may be contaminated 
with pathogenic microorganisms that can cause infectious 
diseases. Improper handling of blood may cause infection to the 
user or other individuals by pathogenic microorganisms.

• This instrument is to be operated by qualified persons only. A 

qualified person is one having adequate knowledge of clinical 
testing and the disposal of infectious waste. Thoroughly read 
this operating manual before use. Anyone who operates the 
instrument for the first time must be assisted by a trained 
person.

• Never touch the strip tray or other parts where the sample

may adhere with your bare hands. During cleaning or 
maintenance of these parts, wear disposable gloves to prevent 
exposure to pathogenic microorganisms.

• Dispose of used samples, tips, test strips, parts and instrument 

in accordance with local regulations for biohazardous waste.

NOTE:

This product is a precision instrument. Handle the instrument with care. 
Do not subject the instrument to strong impact or vibration

© 2019 ARKRAY, Inc.

Introduction

• It is strictly prohibited to copy any part of this operating manual without the 

expressed consent of ARKRAY, Inc.

• The information in this operating manual is subject to change without 

notice.

• ARKRAY, Inc. has made every effort to prepare this operating manual. 

Should you discover anything strange, incorrect or missing, please 
contact your distributor.

This instrument has a caution label on an area that poses a potential 
danger. Please learn about the danger warned of by this label and 
observe the precaution described below.

a

 Strip tray

Do not touch the strip tray with your bare hands. Wear 
disposable gloves to prevent exposure to pathogenic 
microorganisms while setting the test strips or cleaning the 
strip tray.

a

Summary of Contents for Lab 004

Page 1: ... the instrument and the proper procedures for operation and maintenance Follow the instructions in this operating manual in order not to defeat the purpose of the protective features of the instrument For the purchase of reagents consumables or other optional items contact your distributor TAKE THE UTMOST CARE WHEN HANDLING BLOOD This instrument uses blood as sample Blood may be contaminated with ...

Page 2: ...Dual wavelength transmitted light measurement Number of test strips measurable at once 1 Measurement time Listed in package insert of reagent Approximately 4 μL Sample consumption Listed in package insert of reagent 1 sample Approximately 10 minutes in environments 25 C and below Memory capacity 50 measurements Display Segment LCD Flat keypad Micro USB type B 110 W 225 D 54 H mm Weight Approximate...

Page 3: ... battery in place of the provided AC adapter Use a battery with a rating of 5 V and 2 0 2 4 A Do not disassemble the instrument Do not modify the instrument Disassembly and modification of the instrument may result in exposure to pathogenic microorganisms or cause fire or damage to the instrument and consequently lead to personal injury 1 4 2 Precautions for Instrument Relocation Before relocating...

Page 4: ...ior to measurement Do not store a sample in direct sunlight or a location directly exposed to air flow from an air conditioner or other source Ensure the instrument is in a proper environment before turning on the power see 1 4 1 Precautions for Instrument Installation Do not place containers or bottles that contain liquid on the instrument Sample or other liquid that gets inside the instrument ma...

Page 5: ...econds after withdrawing the strip tray 2 2 4 Measuring the control a On the standby screen press the button The control icon will appear b Dispense a drop of the control onto the test strip and then perform the same procedure as when measuring a sample c The measurement results are saved as a control measurement 2 2 5 Understanding results After measurement the following information appears on th...

Page 6: ...ace the parts at recommended intervals 1 Clean the strip tray At the end of the day When measurement for the day is complete be sure to clean the strip tray Prepare cotton swabs distilled water soft cloth alcohol and disposable gloves Use a soft cloth or cotton swab moistened with water or alcohol to clean the strip tray 2 Cleaning the photometric window when E 130 occurs Prepare cleaning pad dist...

Page 7: ...ur distributor 84 07200A Rev OCT 2019 T 999 A problem occurred in communication with an external device Connect the external device with the communication cable correctly E 102 The date and time settings are incorrect Set the date and time correctly E 103 The photometric window is dirty Clean the photometric window E 122 The ambient temperature of the instrument is less than 10 C or over 35 C Adju...

Reviews: