background image

8

General

• The charger converts mains voltage to low DC voltage.
• The charger must only be used for ArjoHuntleigh battery 

type 

NEA

 0100.

• The charger is ready for use as soon as it is plugged in.
• Disconnect the mains by unplugging the charger.
• A green indicator LED indicates the charger is ready for 

charging, there is no On/Off button on the charger.

Charging Intervals

• To prolong life of batteries, charge at regular intervals.
• If the ArjoHuntleigh product is used frequently, it is 

recommended to use two batteries, one in operation and one 
being charged

Charging Procedure (6 Steps)

1 Plug in the charger in a power outlet.

(See Fig. 1)

2 The charger indicator LED turns green when there is no 

battery connected.

3 Attach the battery to the charger

4 Plug the charger connector into the battery. 

5 The charger indicator LED will stay orange during 

charging.

6 Once the battery is charged the LED will turn green. 

Charging the battery can take up to 14 hours depending on the 
battery status (e.g. if the battery is charged for the first time or after 
a longer storage period).

WARNING

Do NOT charge the battery in a sealed 
compartment due to fire hazard.

CAUTION
To avoid breaking the charger when unplugging the 
battery connector: Grip the plug and don’t pull the 
cord.

NOTE
The battery can be left in the charger unit even if it is 
fully charged without being overcharged.

 Fig. 1

Charger
connector

Charging the Battery

Summary of Contents for NEA 1000

Page 1: ...with people in mind INSTRUCTIONS FOR USE NEA 1000 2000 4000 6000 Battery Charger 04 NE 00_4GB February 2015...

Page 2: ...nuous improvement we reserve the right to modify designs without prior notice The content of this publication may not be copied either whole or in part without the consent of ArjoHuntleigh AB WARNING...

Page 3: ...Installation of Charger 7 Charging the Battery 8 Cleaning Instructions 9 Care and Preventive Maintenance 9 Troubleshooting 10 Technical Specifications 11 Labels 13 Electromagnetic Compatibility 14 Par...

Page 4: ...ll not be held responsible for any accidents incidents or lack of performance that occur as a result of any unauthorized modification to its products Service and Support A service routine has to be do...

Page 5: ...for charging ArjoHuntleigh battery NEA 0100 only in accordance with the in structions outlined in the Instructions for Use IFU All other use must be avoided Installation Service Requirements The expec...

Page 6: ...d explosion or fire never use the equipment in oxygen rich environment in presence of heat source or flammable anaesthetic gases Battery and Charger Safety Practices Be careful not to drop the battery...

Page 7: ...rough inspection for damage 2 If any part is missing or damaged do NOT use the product NOTE If you have any questions please contact your local ArjoHuntleigh representative for support and service The...

Page 8: ...n a power outlet See Fig 1 2 The charger indicator LED turns green when there is no battery connected 3 Attach the battery to the charger 4 Plug the charger connector into the battery 5 The charger in...

Page 9: ...reventive Maintenance Schedule WARNING To avoid malfunction resulting in injury make sure to conduct regular inspections and follow the recommended maintenance schedule In some cases due to heavy use...

Page 10: ...injury and or unsafe product the maintenance activities must be carried out at the correct frequency by qualified personnel using correct tools parts and knowledge of procedure Qualified personnel mus...

Page 11: ...ttery charger weight 0 15 kg 0 33 lbs Flammability standard UL94V 1 Ambient temperature 10 C to 40 C 50 F to 104 F Operation 30 C to 50 C 22 F to 122 F Transport Storage Relative humidity 30 75 Operat...

Page 12: ...o indicate that it is electronic equipment covered by the Directive 2012 19 EU on waste electrical and electronic equipment In European countries the crossed out wheeled bin WEEE symbol reminds you th...

Page 13: ...max IP2X Protection class The product must be used in a dry location only Separate electrical and electronic components for recycling in accordance with the European Directive 2012 19 EU WEEE Insulat...

Page 14: ...Guidance and Manufacturer s Declaration Electromagnetic Emissions The Battery Charger is intended for use in the electromagnetic environment specified below The customer or the user of the Battery Ch...

Page 15: ...should be that of a typical commercial or hospital environment Surge IEC 61000 4 5 1 kV line s to line s 2 kV line s to earth 1 kV line s to line s 2 kV line s to earth Mains power quality should be t...

Page 16: ...er and d is the recommended separation distance in metres m Field strengths from fixed RF transmitters as determined by an electromagnetic site survey a should be less than the compliance level in eac...

Page 17: ...17 Battery Charger NEAX000 XX Extra battery NEA 0100 24 V Battery holder 8350891 031 Parts and Accessories...

Page 18: ...18 Intentionally left blank...

Page 19: ...E mail international arjohuntleigh com ITALIA ArjoHuntleigh S p A Via di Tor Vergata 432 IT 00133 ROMA Tel 39 0 6 87426211 Fax 39 0 6 87426222 E mail Italy promo arjohuntleigh com NEDERLAND ArjoHuntle...

Page 20: ...oHuntleigh GETINGE and MAQUET ArjoHuntleigh focuses on patient mobility and wound management solutions GETINGE provides solutions for infection control within healthcare and contamination prevention w...