13
11. Technical Specification
If the pump is stored in conditions outside of the
“Operating” ranges, it should be allowed time to
stabilise at normal operating conditions before use.
Pump
Model:
Flowtron
Excel
Part Numbers:
UK/AUS
KSA
USA
247003 - UK
247003AU - AUS
247004 (247EUR) - EU
247009KSA
247001
Supply Voltage:
230 V
230 V
120 V
Supply Frequency:
50 Hz
60 Hz
60 Hz
Power Input:
35 VA MAX
35 VA MAX
35 VA MAX
Size:
133 x 152 x 275 mm
Weight:
2.7 kg
Case Material:
Fire Retardant ABS Plastic
Plug Fuse Rating:
5A BS1362 (UK ONLY)
Pump Fuse Rating:
T1AL 250 V
Degree of protection
against electric shock:
Class II, Double Insulated with Functional Earth
Type BF
Degree of protection
against liquid ingress:
IPX0 - No protection
Mode of operation:
Continuous
Cycle or Therapy Modes:
60 seconds total
12 seconds inflation
48 seconds deflation
Pressure Range:
30 - 60 mmHg (±4 mmHg)
ENVIRONMENTAL INFORMATION
Condition
Temperature Range
Relative Humidity
Atmospheric Pressure
Operating
+10 °C to +40 °C
(+50 °F to +104 °F)
30% to 75%
(non-condensing)
700 hPa to 1060 hPa
Storage and Transport
(Long Term)
+10 °C to +40 °C
(+50 °F to +104 °F)
20% to 95%
(non-condensing)
700 hPa to 1060 hPa
Storage and Transport
(Short Term)
-20 °C to +50 °C
(-4 °F to +122 °F)
20% to 95%
(non-condensing)
500 hPa to 1060 hPa
Products should be well packaged and stored in a well-ventilated, dry and non-corrosive environment.