33
746-449-SA-5
The zero point can only be set if the load on the bed is 25kg or
less above the factory calibrated zero.
If the user attempts to set the zero point more than once,
the display will show error code
E001
. To clear the error,
disconnect the bed from the electricity supply for ten
seconds; when power is restored, the zero point will revert
to the previous setting.
Weighing
Press the
Enable
key followed by the
Weigh
key.
The display will show a moving circular pattern until a
stable weight reading is obtained.
The patient’s weight will be displayed for ten seconds
then the display will go blank.
The weight is displayed rounded to the nearest 500g; this
can be temporarily changed to 100g by pressing the
Select
Weight Display
key. The display will revert to 500g next
time the
Weigh
key is pressed.
The 100g weight display should only be used for comparative
readings.
Auto
compensation
The Auto compensation facility allows weight to be added
to or removed from the bed, without affecting the
indicated patient weight.
To use Auto compensation:
With the patient lying on the bed, press the
Enable
key
followed by the
Auto Compensation
key.
The display will indicate
AUTO
.
Add or remove accessories, bed linen, etc. as required.
Press the
Enable
key followed by the
Auto
Compensation
key. The indicators above the keys will go
out and the display will show the patient’s weight as
described above.