10
746-449-SA-5
Caution
Do not allow the electrical supply cable to touch the bed frame
as this may affect the accuracy of the weighing system.
An equipotential terminal (D) is located at the head end of
the bed, above the roller buffer on the patient’s right side.
When other electrical equipment is within patient / user
reach, the potential difference between this equipment can
be equalised by interconnecting the equipotential
terminals of each piece of equipment.
Caution
Before the first use, or if the bed has been unused for more than
three months, connect the bed to the electricity supply for 24
hours to recharge the backup battery. Failure to do this may
reduce the life of the battery.
Model 9000D beds should be positioned with the roller buffers at
least 20cm from the rear wall to prevent the head end panel
striking the wall when the bed is in head down tilt
(Trendelenburg).
Recommended positioning for Model 9000D