746-577-UK-6
59
Recommended separation distances between portable and mobile RF
communications equipment and the bed
The bed is intended for use in an electromagnetic environment in which radiated RF
disturbances are controlled. The customer or the user of the bed can help prevent
electromagnetic interference by maintaining a minimum distance between portable and
mobile RF communications equipment (transmitters) and the bed as recommended below,
according to the maximum output power of the communications equipment.
Rated maximum output
power of transmitter
W
Separation distance according to frequency of transmitter
m
150kHz to 80MHz 80MHz to 800MHz 800MHz to 2.5GHz
d
P
=
1 2
.
d
P
=
1 2
.
d
P
=
2.3
0.01
0.12
0.12
0.23
0.1
0.38
0.38
0.73
1
1.2
1.2
2.3
10
3.8
3.8
7.3
100
12
12
23
For transmitters rated at a maximum output power not listed above, the recommended separation distance
d
in
metres (m) can be estimated using the equation applicable to the power of the transmitter, where
P
is the
maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer.
NOTE 1 At 80MHz and 800MHz, the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by
absorption and reflection from structures, objects and people.
Summary of Contents for E5X
Page 1: ...EN CN 746 577 CN_05 10 2018 ENTERPRISE E5000X E5X Փ 䇈ᯢк Instructions for Use ...
Page 61: ...746 577 UK 6 60 This Page Intentionally Left Blank ...
Page 62: ...746 577 UK 6 61 This Page Intentionally Left Blank ...
Page 83: ...746 577 CN_05 82 抓住安装点旁 2 的侧护栏并将侧护栏从床架上拉开 3 4长度的护栏的拆卸 安装侧护栏时 需要将护栏支架移至床架安装点 2 并用 安全销固定住 2 ...
Page 121: ...746 577 CN_05 120 本页特意留空 ...
Page 122: ...746 577 CN_05 121 本页特意留空 ...