746-577-UK-6
12
Contra-
indications
The safe working load of the bed is
250kg.
The maximum recommended
patient weight is 185kg.
The safe working load is calculated as follows (in
accordance with EN 60601-2-52:2010):
Maximum patient weight
185kg
Mattress
20kg
Accessories (including attached loads)
45kg
TOTAL
250kg
WARNING
If the combined weight of the mattress and accessories exceeds
65kg, the maximum patient weight must be reduced accordingly.
The recommended patient height
is between 146cm and 190cm.
At the discretion of the carer, patients taller than 190cm
may be accommodated by extending the bed - refer to
“Bed length adjustment” in Chapter 4. Ensure that the
patient’s height does not exceed the “In-bed length”
shown in Chapter 7.