AKRON STREAMLINE Model No.’s:- 2221, 2221A 2221B, 2221C, 2222, 2222A, 2222B, 2222C, 2223, 2321, 2322, 2323
2231, 2231A, 2231B, 2231C, 2232, 2232A, 2232B, 2232C, 2331, 2332
11
6. Cleaning and Decontamination
Frequency of routine couch cleaning will depend on exact application.
CLEANING / DECONTAMINATION
Locate couch in a suitable cleaning area with castors disengaged or wheels locked.
1.
Wear protective clothing: plastic disposable apron and gloves.
2.
Disconnect any mains supply and lock couch wheels.
3.
Control handsets can be wiped over using a 70% Isopropranol alcohol wipe.
4.
Prepare a solution of warm water and neutral detergent.
5.
Start at the top of the couch and work your way downwards.
6.
Wipe over head end working towards mid torso.
7.
Wipe over remaining horizontal patient surface, working towards foot end.
8.
Wipe all horizontal parts under the patient surface.
9.
Wipe all vertical parts under the patient surface.
10.
Wipe over shroud (if fitted).
11.
Wipe over all parts of wheels.
12.
Ensure all parts of the couch are dry before re-use.
13.
Safely discard cleaning materials.
14.
Wash hands thoroughly.
If DECONTAMINATING then repeat steps 1) to 14) using a dilute Hypochlorite / warm water
mixture (10,000ppm) instead of warm water and soap.
Put couch back into service.
WARNING
Do not use Alkalis or other organic solvents for routine cleaning.
Replace torn or damaged vinyl immediately.
Always cover vinyl with fresh paper or sheet between patients.
If contamination is known or suspected always decontaminate couch
before use.
Do not saturate hinge pin location and it is essential to ensure that they
are dry before re-use.
Disconnect any mains supply.