background image

5

Intended Use

This equipment must be used in line with its 
intended use and in accordance with these 
safety instructions. Anyone using this 
equipment must also have read and 
understood the instructions in this 

Instructions for Use (IFU)

.

If there is anything you are not sure about, 
contact your Arjo representative.

This charger is intended for charging Arjo battery 

NEA 

0100

 only, in accordance with the instructions outlined 

in the 

Instructions for Use

.

All other use must be avoided.

Life time of the equipment

The expected life of the 

Battery Charger NEA8000-XX

is unless otherwise is stated ten (10) years, if subject to 
preventive maintenance being carried out in accordance 
with the instruction for 

Care and Preventive 

Maintenance

 found in the 

Instructions for Use (IFU)

.

Assembly of parts must be mounted according to the 
instructions in the 

Instructions for Use (IFU)

.

Summary of Contents for NEA 8000

Page 1: ...NEA 8000 04 NE 08_8EN 11 2019 INSTRUCTIONS FOR USE Battery Charger ...

Page 2: ...us improvement we reserve the right to modify designs without prior notice The content of this publication may not be copied either whole or in part without the consent of Arjo WARNING To avoid injury always read this Instruction for Use and accompanying documents before using the product Mandatory to read the Instructions for Use ...

Page 3: ... 7 Installation of Charger 7 Charging the Battery 8 Cleaning Instructions 9 Care and Preventive Maintenance 9 Troubleshooting 10 Technical Specifications 11 Approval List of Standards 12 Labels 13 Electromagnetic Compatibility 14 Parts and Accessories 17 Contents ...

Page 4: ...not be held responsible for any accidents incidents or lack of performance that occur as a result of any unauthorized modification to its products Service and Support A service routine has to be done on your Battery Charger byArjo authorised service personnel to ensure the safety and operating procedures of your product See section Care and Preventive Maintenance on page 9 If you require further i...

Page 5: ...tended for charging Arjo battery NEA 0100 only in accordance with the instructions outlined in the Instructions for Use All other use must be avoided Life time of the equipment The expected life of the Battery Charger NEA8000 XX is unless otherwise is stated ten 10 years if subject to preventive maintenance being carried out in accordance with the instruction for Care and Preventive Maintenance fo...

Page 6: ...id explosion or fire never use the equipment in oxygen rich environment in presence of heat source or flammable anaesthetic gases Battery and Charger Safety Practices Be careful not to drop the battery Contact the appropriate local authority for advice when disposing of batteries Check label on the charger and battery WARNING To avoid electric shock make sure that the equipment is connected to Con...

Page 7: ...nt emergency Actions Before Every Use 2 Steps 1 Carry out a thorough inspection for damage 2 If any part is missing or damaged Do NOT use the product NOTE If you have any questions please contact your local Arjo representative for support and service The contact information is listed at the end of this IFU Installation of Charger Installation Keep the Battery Charger away from main access areas Th...

Page 8: ...er in a power outlet 2 Plug the charger connector into the battery See Fig 1 3 Attach the battery to the charger 4 The charger LED indications Step 1 Orange LED for fast charge Step 2 Green LED with short Yellow light intervals until final end charge Step 3 Solid green LED indicating a fully charged battery Charging the battery can take up to 14 hours depending on the battery status e g if the bat...

Page 9: ...resulting in injury make sure to conduct regular inspections and follow the recommended maintenance schedule In some cases due to heavy use of the product and exposure to aggressive environment more frequent inspections should be carried out Local regulations and standards may be more stringent than the recommended maintenance schedule CAREGIVER OBLIGATIONS Action Check Every WEEK Visually check a...

Page 10: ...to repair it Troubleshooting WARNING To avoid injury and or unsafe product the maintenance activities must be carried out at the correct frequency by qualified personnel using correct tools parts and knowledge of procedure Qualified personnel must have documented training in maintenance of this device QUALIFIED PERSONNEL Action Check Every YEAR Check battery and Battery Charger X Check all vital p...

Page 11: ...NEA 0100 24V 1 4 kg 3 08 lbs Battery charger weight 0 15 kg 0 33 lbs Flammability standard UL94V 1 Ambient temperature 10 C to 40 C 50 F to 104 F Operation 30 C to 50 C 22 F to 122 F Transport Storage Relative humidity 30 75 Operation 10 80 Transport Storage Atmospheric pressure 800 1060 hPa Operation 500 1060 hPa Transport Storage Power Cord Supply Not allowed to be replaced Screw for mounting th...

Page 12: ... electronic products batteries and accumulators must be taken to separate collection at end of their working life This requirement applies in the European Union Do not dispose these products as unsorted municipal waste You can return your device and accessories to Arjo or you can contact your local authorities for advice on disposal Follow local regulations Approval List of Standards Certificate S...

Page 13: ...er consumption max IP2X Protection class The product must be used in a dry location only Insulation class Class II Separate electrical and electronic components for recycling in accordance with the European Directive 2012 19 EU WEEE Read the IFU before use PSE Product Safety of Electrical Appliance Materials The equipment has been certified for Japan SE XXXX XXXX Serial Number CE marking indicatin...

Page 14: ...on Guidance and Manufacturer s Declaration Electromagnetic Emissions The Battery Charger is intended for use in the electromagnetic environment specified below The customer or the user of the Battery Charger should assure that it is used in such an environment Emissions Test Compliance Electromagnetic Environment Guidance RF emissions CISPR 11 Class B The Battery Charger uses RF energy only for it...

Page 15: ... should be that of a typical commercial or hospital environment Surge IEC 61000 4 5 1 kV line s to line s 2 kV line s to earth 1 kV line s to line s 2 kV line s to earth Mains power quality should be that of a typical commercial or hospital environment Voltage dips short interruptions and voltage variations on power supply input lines IEC 61000 4 11 5 UT 95 dip in UT for 0 5 cycle 40 UT 60 dip in ...

Page 16: ...the recommended separation distance in metres m Field strengths from fixed RF transmitters as determined by an electromagnetic site survey a should be less than the compliance level in each frequency range b Interference may occur in the vicinity of equipment marked with the following symbol Radiated RF IEC 61000 4 3 3 V m 80 MHz to 2 5 GHz 3 V m a Field strengths from fixed transmitters such as b...

Page 17: ...ry holder 8350891 031 Extra Battery NEA 0100 24V Battery Charger NEA 8000 WW delivered with interchangeable mains plug EU US UK NEA 8000 AU delivered with interchangeable mains plug AU JP Parts and Accessories ...

Page 18: ...18 Intentionally left blank ...

Page 19: ...ail Italy promo arjo com MIDDLE EAST Arjo Middle East FZ LLC Of ce 908 9th Floor HQ Building North Tower Dubai Science Park Al Barsha South P O Box 11488 Dubai United Arab Emirates Direct 971 487 48053 Fax 971 487 48072 Email Info ME arjo com NEDERLAND Arjo BV Biezenwei 21 4004 MB TIEL Postbus 6116 4000 HC TIEL Tel 31 0 344 64 08 00 Fax 31 0 344 64 08 85 E mail info nl arjo com NEW ZEALAND Arjo Lt...

Page 20: ...re ergonomic patient handling personal hygiene disinfection diagnostics and the effective prevention of pressure ulcers and venous thromboembolism we help professionals across care environments to continually raise WKH VWDQGDUG RI VDIH DQG GLJQL HG FDUH YHU WKLQJ ZH GR ZH GR ZLWK SHRSOH LQ PLQG ArjoHuntleigh AB Hans Michelsensgatan 10 211 20 Malmö Sweden www arjo com ...

Reviews: