21
SYMBOLS USED
Important Operational
Information
Warning of possible hazard to system,
patient or staff
Possible electrical shock hazard
EMC
The MaxxAir ETS (Expandable Turning Surface) Mattress Replacement Systems’s Air
Control Unit uses RF energy only for its internal function. Therefore, its RF emissions
are very low and are not likely to cause interference in nearby electronic equipment. The
MaxxAir ETS was tested to the following standards:
• IEC 60601-1-2: Second Edition
• EN 5501: 1998 Group 1, Class B
• EN 61000-3-2: Class A
• EN 61000-3-3
CLASSIFICATION
Classification in accordance with UL 60601-1, CSA C22.2 No. 601.1, and IEC 60601-1.
• Class II
• Type BF
• IPX1
•
No Sterilization
• Not for Use with Flammable Anesthetic Mixture with Air, Oxygen or Nitrous Oxide
• Continuous Operation
• Ordinary Equipment
ENVIRONMENTAL STORAGE / TRANSIT CONDITIONS
Temperature Range: -20 to 60° C (-4 to 140° F)
Relative Humidity Range: 30 to 95%
Atmospheric Pressure Range: 500 to 1060 hPa
ENVIRONMENTAL OPERATING CONDITIONS
Temperature Range: 10 to 40° C (50 to 104° F)
Relative Humidity Range: 30 to 95%
Atmospheric Pressure Range: 500 to 1060 hPa