49
Preventive maintenance
This product is subject to wear and tear during use. To ensure that it
continues to perform within its original specification, preventive
maintenance procedures should be carried out at the intervals shown.
If the result of any of these tests is unsatisfactory, contact Arjo or an
approved service agent.
WARNING
This list indicates the minimum recommended level of preventive maintenance. More
frequent inspections should be carried out when the product is subjected to heavy
use or aggressive environments, or where required by local regulations.
Failure to carry out these checks, or continuing to use the product if a fault is found,
may compromise the safety of both the patient and caregiver. Preventive maintenance
can help to prevent accidents.
NOTE
Product cannot be maintained and serviced while in use with the patient.
Actions to be done by caregiver
Daily
Weekly
Check operation of split side rails
9
Visually check castors
9
Check operation of the manual CPR release handles on both
sides of the bed
9
Visually check power supply cord and mains plug
9
Carry out a full test of all electrical bed positioning functions
(backrest, height, tilt, etc.)
9
Check that the patient controls, caregiver controls and Attendant
Control Panels operate correctly
9
Check that the weighing system controls operate correctly
9
Check operation of the anti-entrapment system and clean the
sensor lenses (see page 46) (Optional)
9
Check the mattress for damage and fluid ingress
9
Examine the lifting pole, strap and handle (Optional)
9