Clean and disinfect Hydromassage System with Automatic Disinfection
This procedure is a complement when cleaning the bath system.
1. Check the level of disinfectant.
2. Open the drain.
3. Remove the suction inlet screen and clean it separately. See Figure 1.
4. For treatment hose only
: pull out the Hydromassage system nozzle and
clean it separately. See Figure 2.
5. For treatment hose only:
Insert the treatment hose in the Hydromassage
system outlet. See Figure 3.
6. Press the Disinfection button on the control panel to start the automatic
GLVLQIHFWLRQF\FOH'LVLQIHFWDQWLVÀXVKHGWKU\GURPDVVDJHV\VWHP
nozzle/treatment hose. The disinfectant cycle ends with the Hydromassage
system being rinsed with water.
7. For treatment hose only
: When the disinfectant cycle is completed, remove
the treatment hose and let it hang dry on the holder. Place the Hydromassage
system nozzle back into the Hydromassage system outlet.
8. Open the access door on the left side of the panel and lift out the disinfection
spray handle or use a disinfection spray bottle.
9. For disinfection spray handle only
: Direct the disinfectant spray towards
the bottom of the bathtub and press the Disinfection button on the control
panel.
10.
6SUD\GLVLQIHFWDQWLQWRWKHVXUIDFHRYHUÀRZRXWOHW6HH)LJXUH
11. For disinfection spray handle only
: Press the Disinfection button on the
control panel to turn it off, and place the disinfectant spray handle back
behind the access door.
12. Allow appropriate contact time according to the instructions of the
disinfectant.
13. When disinfection is completed, reassemble the suction inlet screen.
14. Press the Patient Shower button on the control panel.
15. Rinse all treated parts clean with plenty of water having a temperature of
approximately 25 °C (77 °F) to remove disinfectant.
Continue with remaining disinfection instructions according to section
Disinfect (step 27-42 on page 1).