Clean and disinfect Hydromassage System with Automatic Disinfection

This procedure is a complement when cleaning the bath system.

1.  Check the level of disinfectant.

2.  Open the drain.

3.  Remove the suction inlet screen and clean it separately. See Figure 1.

4.  For treatment hose only

: pull out the Hydromassage system nozzle and 

clean it separately. See Figure 2. 

5.  For treatment hose only:

 Insert the treatment hose in the Hydromassage 

system outlet. See Figure 3.

6.  Press the Disinfection button on the control panel to start the automatic 

GLVLQIHFWLRQF\FOH'LVLQIHFWDQWLVÀXVKHGWKU\GURPDVVDJHV\VWHP

nozzle/treatment hose. The disinfectant cycle ends with the Hydromassage 
system being rinsed with water.

7.  For treatment hose only

: When the disinfectant cycle is completed, remove 

the treatment hose and let it hang dry on the holder. Place the Hydromassage 
system nozzle back into the Hydromassage system outlet. 

8.  Open the access door on the left side of the panel and lift out the disinfection 

spray handle or use a disinfection spray bottle.

9.  For disinfection spray handle only

: Direct the disinfectant spray towards 

the bottom of the bathtub and press the Disinfection button on the control 
panel. 

10. 

6SUD\GLVLQIHFWDQWLQWRWKHVXUIDFHRYHUÀRZRXWOHW6HH)LJXUH

11. For disinfection spray handle only

: Press the Disinfection button on the 

control panel to turn it off, and place the disinfectant spray handle back 
behind the access door. 

12. Allow appropriate contact time according to the instructions of the 

disinfectant.

13. When disinfection is completed, reassemble the suction inlet screen.

14. Press the Patient Shower button on the control panel.

15. Rinse all treated parts clean with plenty of water having a temperature of 

approximately 25 °C (77 °F) to remove disinfectant.

Continue with remaining disinfection instructions according to section 
Disinfect (step 27-42 on page 1).