23
Technical Data
Measurements, Volume, Weight
Weight
14 kg (31 lbs) without flush
22 kg (49 lbs) with flush tank
14 kg (31 lbs) with flush valve
Width
455 mm (17 7/8”) without flush
750 mm (29 1/2”) with flush tank
455 mm (17 7/8”) with flush valve
Height
510 mm (20”)
Depth
190 mm (7 1/2”)
Thermometer
Range
20-60 °C (68-140 °F)
Accuracy
+/-2 °C (3.6 °F)
Resolution
0.5 °C (0.9 °F)
Water supply requirements
Cold water
G3/4” male (customers connection)
Warm water
G3/4” male (customers connection)
Connect with shut off valves, 3/4” connection
Maximum static pressure
6 bar (600 kPa, 87 psi)
Minimum operating pressure with disinfection
Minimum operating pressure without disinfection
Minimum operating pressure (for BCA122X-04)
2 bar (200 kPa, 29 psi)
0.8 (80 kPa,12 psi)
2 bar (200 kPa, 29 psi)
Minimum flow
20 l/minute (5.3 US gallon/minute)
Recommended cold water temperature
Recommended hot water temperature
2 °C - 20 °C (36 °F - 68 °F)
60 °C - 80 °C (140 °F - 176 °F)
Water drainage requirements
Minimum drainage capacity - Shower & Disinfection Panel
25 l/minute (6.6 US gallon/minute)
Minimum drainage capacity - Flush sinks
100 l/minute (26.4 US gallon/minute)
DIV 110
Technical Specifications