10
AtmosAir
Pump Installation
Pump installation steps for A and AR models:
1.
Place pump on a solid stationary surface or suspend on end of bed frame with
built-in hanger.
2.
Attach pump hoses to mattress, as shown at right (Figure 7):
A Models (alternating pressure):
• Attach
blue
hoses to the mattress connectors on the
patient left side
(marked with
BLUE tag
) of the mattress for
alternating pressure
.
AR Models (alternating pressure / rotation):
• Attach
blue
hoses to the mattress connectors on the
patient left side
(marked with
BLUE tag
) of the mattress for
alternating pressure
.
• Attach
red
hoses to the mattress connectors on the
patient right side
(marked with
RED tag
) of the mattress for
rotation
.
3.
Ensure air hoses are not kinked and will not be pinched by any articulated bed
mechanisms.
4.
Attach hose connectors to the pump.
5.
Plug pump unit into a properly-grounded wall outlet.
6.
Verify power to this outlet is not controlled by a wall switch.
7.
Switch pump on.
Pump Hoses
Mattress
Connectors
Tag
Pump Hoses
Mattress
Connectors
Tag
Figure 7 120V
Figure 7a 230V