Changes in Patient Position

When the patient changes position from lying to sitting, or reversed,

LWPD\EHQHFHVVDU\LQRUGHUWRPD[LPLVHWKHEHQH¿WRIWKHVXSSRUW

surface, to make adjustments to the mattress pressure.

From Lying to Sitting

 – increase pressure control.

From Sitting to Lying

 – decrease pressure control.

This adjustment should be in conjunction with independent clinical 
determination of appropriate support.

Manual-only system

Convert to manual-only system:

Press and hold the minus (-) button at least 10 seconds in unlock 
mode.

Return to semi-automatic system:

Press and hold the plus (+) button at least 10 seconds in unlock 
mode.

How to unplug the heel zone valves:

There are 5 Heel Zone cells located in the 16 Active (Alternating) 

FHOOVWKDWFDQEHVHOHFWLYHO\DQGSHUPDQHQWO\GHÀDWHG

Press the button (1) on the valve and pull on the tube (2) to separate
the two parts in order to disconnect and permanently vent the upper 

SDUWRIDKHHOFHOOWRSURYLGHKHHORႉRDGLQJ

See Figure. 3.

When this function is no longer required then reconnect the two parts

RIWKHYDOYHWRUHLQÀDWHWKHFHOO

Figure 3

Transport Mode

Transporting the patient on the Alpha Active 5 mattress overlay.
1.

Disconnect the tube-set from the pump. See Figure 4.

2.

Place the cap over the end of the tubeset. See Figure 5.

With the cap attached the mattress will automatically switch to
transport mode.
Mattress support will remain for up to 12 hours in Transport Mode.
To resume normal operation, reconnect the tubeset and run the pump.

Figure 4

Figure 5

NOTE

$PD[LPXPRIWZRKHHOFHOOVDUHDOORZHGWREHPDQXDOO\
GHÀDWHGDWDQ\WLPH

CAUTION

7UDQVSRUWPRGHWHPSRUDULO\RႇHUVVXSSRUWIRUXSWR
KRXUV,WLVUHFRPPHQGHGWKDWZKHQLQWUDQVSRUWPRGH
WKHSDWLHQWLVIUHTXHQWO\PRQLWRUHG

® and ™ are trademarks belonging to the Arjo group of companies.
© Arjo 2022.
As our policy is one of continuous improvement, we reserve the right to modify designs without prior notice. The
content of this publication may not be copied either whole or in part without the consent of Arjo.

RAPID DEFLATE

,17+((9(172)&$5',$&$55(67

In the event of that CPR needs to be administered:

7RDFWLYDWH5DSLG'HÀDWH

:

3XOOWKHUHGVWUDSODEHOOHG&35IRUWKHPDWWUHVVWRGHÀDWH

7RUHVHW5DSLG'HÀDWH

Replace the stopper securely and reconnect the air tubes 
to the pump.