Bair Paws
®
Model 850 Warming Unit Operator’s Manual
10
General Maintenance
Calibrating the Operating Temperatures
Operating temperature calibration should be verified every 6 months and after performing service
procedures. In addition, if the ambient temperature where the warming unit will be used is higher
than 24°C (75°F), the unit should be recalibrated before it is placed in service. You will need a Bair
Paws Model 90055 temperature test kit to perform the calibration. Calibration instructions are
included with the test kit.
Replacing the Filter
Replace the filter every 6 months or sooner if necessary. To order a replacement filter, contact Arizant
Healthcare customer service using either of the telephone numbers listed on page 11. Instructions for
replacing the filter will be included with the replacement filter.
Cleaning the Warming Unit
Warning:
Do not immerse any part of the warming unit while cleaning it. Moisture will damage the
components, and thermal injury may result.
Precaution
Do not use a dripping wet cloth to clean the warming unit. Moisture may seep into the electrical
•
contacts and damage the components.
Do not use harsh solvents to clean the warming unit.* Solvents may damage the labels and other
•
plastic parts.
M
ethod
Disconnect the warming unit from the power source before cleaning.
1.
Wipe the cabinet, the temperature controller, and the outside of the hose with a damp, soft cloth
2.
and a mild cleaning solution or antimicrobial spray.
Dry with a separate soft cloth.
3.
* WINDEX® multi-use liquid cleaner, ENVIROCIDE® disinfectant, and ALCONOX® industrial cleaner are three products that
can be used to clean the equipment, as well as 3% household bleach, isopropanol, ethanol, ammonia or other phosphate-based
glass cleaning solutions.
WINDEX is a registered trademark of S.C. Johnson & Son, Inc.
ENVIROCIDE is a registered trademark of Metrex Research Corporation.
ALCONOX is a registered trademark of Alconox, Inc.