© ArianaSpas
17 |
P a g e
o
The suction force of vacuum&RF should not be too high when operating the on the face, so as to avoid
scrapping.
o
During operation, the laser plate must be fixed on the customer first, and then the instrument should
be started to adjust the energy. The energy should be adjusted slowly from low to high to make the
customer feel warm and comfortable.
o
Must apply the product that of higher oil content, and keep it moist.
o
The laser plate can also be used alone, directly fixed in the fat part that needs to lose weight and used
after adjusting the energy. The operation time of one part is about 20 minutes. Combine massage, RF
and ultrasonic treatments for better results.
o
During operation, the whole surface of the laser plate should not only touch a small area but be in full
contact with the skin
8. Troubleshooting & Solutions
1.
The device cannot be started, and the light on the back of the device is not working?
A.
Make sure the power cord is connected to a good power socket.
B.
Check the fuse tube on the back of the device. Maybe
it’s
loose or burnt out.
2.
The RF output is not working?
A.
Please check the handle of the device and the plug. Whether they are closely connected with each
other.
B.
Please check whether the treatment area has been cleaned. Grease or oil essential products may
cause poor contact between handle and human body, which will result in no RF output.
3.
Weakened RF output?
A.
Please check whether the handle and the plug is closely connected.
B.
Please check whether there is non-conductive grease on the handle, which will cause poor contact
and weaken theoutput.
C.
Please check whether the product used is the adaptive product specified by the device.
4.
No suction or with small suction?
A.
Turn off the instrument and check the specialized filter, which may need to be replaced.
B.
Check the rubber ring of the handle plug since a worn rubber ring may cause air
leakage.
C.
Check whether the oil filter cup outside the filter is tightened, and whether the rubber ring is
worn.
It’s
possible that air leakage in this palce may lead to poor air pressure.
D.
If the above-mentioned solutions are not working, please contact instrument dealer for
assistance.
5.
The device can be started, but the monitor shows wrong information?
A.
Take out the plug in the back of the instrument, and wait for about 1 minute and re-plug and restart
the device.
B.
If the above-mentioned solutions are not working, please contact instrument dealer for
assistance.