MAINTENANCE AND FAULT FINDING
14.147.B_QUICK100M en.SM.V1.00.doc
ARGUS Medical AG 11.08.04 / PJ
17 / 22
7. MAINTENANCE AND FAULT FINDING
7.1. Maintenance
CAUTION!
Only authorized persons who have been instructed by ARGUS Medical AG or by
the trained local distributor are allowed to service the QUICK
®
100 M docking sta-
tion. In case of repair request, send the unit to the local distributor with a report
outlining the exact nature of the failure. More information is available from:
ARGUS Medical AG
CH-3627 Heimberg / Switzerland
E-mail:
info@argusmedical.com
CAUTION!
The safety standard check has to be performed at least every 24 month or after
10'000 hours of operation. The check has to be done in accordance to the chapter
11.
No special maintenance of the QUICK
®
100 M docking station is necessary. There
are no wear and tear parts.
Interferences through external high frequency electromagnetic fields (e.g. in com-
bination with surgical equipment) are unknown. In case of doubt we suggest that
you contact your local distributor.
7.2. Fault code
The fault code in case of a technical failure is displayed in the history file, refer to
chapter 5. The possible fault codes registered in the history event log file are listed
in the table below:
Fault Failure:
Code:
21
ROM
test
22
ROM
check
(Runtime)
23
RAM
test/check
24
XRAM
test/check
25
CPU
test
26
Invalid state (in PC-Configuration mode)
27
Eeprom
data
invalid
32
5Volt supply out of range
33
20Volt supply out of range (delayed 5s)
44
Address invalid for config-eeprom
45
Address invalid for history-eeprom
Exceptionally a fault code not listed above may appear. In this case we recom-
mend changing the main board.
7.3. Specifications
Please refer to the user manual for the specifications (cap. 7).