RE ENR.101 – Rèv E – Le 08/04/2014
page : 11
> Hydraulic diagram
>
Pressure mounted
> Observe the maximum operating pressures
of the pressure generator
Maximum operating pressure : CEH-800B
820 bar and CEH-1250B
1270 bar
CEO-700B
720 bar
a)
Connect the instrument on the connector
4
(make sure that o-ring are well positioned).
b)
Check the valve is open and the capstan is in end position stop (turn in the opposite direction
of clockwise).
c)
Close the valve
7
.
d)
Turn the capstan
1
in the direction of clockwise until reaching the calibration pressure.
Maximun operating pressure :
•
CEH-800B (oil)
820 bar
•
CEH-1250B (oil)
1270 bar
•
CEO-700B (water)
720 bar
e)
Realize the measurement point.
>
Reduce pressure
f)
After the readings measures, turn the capstan
1
in the opposite direction clockwise
g)
Open the valve
7
.
h)
The instruments can be disassembled.
Chap.5 - Cleaning/Maintenance
a) Cleaning
The generator cleaning and checking fluid levels are the only periodic maintenance required. No
additional maintenance should be necessary if all instructions are followed.