NAME

4.8 Reading of the configuration

   

                                   Reading of the input parameters

   

                                     Reading of the measure display parameters

                                           Reading of the analog output parameters

                                           

(option analog output)

                                            Reading of the communication parameters

                                           

(option digital output)

                                            Reading of the alarm parameters

                                           

(option 2 relays)

                                            Reading of the safety parameters for sensor  

                                            rupture or self-diagnosis of the outputs
                                            Reading of the display programming parameters  

                                           
                                            Reading of the configuration file name 
                                            Reading of the instrument internal parameters

In each reading submenu, use keys 

and 

to move, and key 

to 

visualise the parameters.
If no key is pressed during 20 s., the instrument will revert to measure display. 

During the reading of the programming, the TPI-SI will carry on measuring and 

managing its output options.

Validation / Vertical move

REL

SECU

INFO

NULL

NULL

YES

=

NULL

NULL

YES

NULL

NO

=

=

NULL

NO

=

YES

=

INFO

TPI-SI

31
AN

PROG

xxxx

ORIG

xx.xx

CSUM

PC-x

65F2

Validation / Vertical move

name of the instrument

(1)

F: fast version

A: analog output

R: relay outputs

N: digital data link

Submenu

type of instrument
available options (1)

identification number of the instrument

programme version number

origin of the instrument

check sum of the programme

no

4.9  Access code

An access code adjustable from 0000 to 9999 serves to protect the converter and its 
setpoints from unauthorized programming, and to lock the access to some functions.

  0 0 0 0       Factory code

    x  x  x   x

                 0 to 5        Access to the display shifting

                 6 to 9        No access

            0 to 5            Access to the display and output simulations 

            6 to 9            No access

        0 to 5                 Access to the teleloading of files

        6 to 9                 No access

    0 to 5                     Access to the quick entering of alarm setpoints

    6 to 9                     No access

4.10     Programming of a new access code  

 

       with the µconsole

PCOD

OLD

NO

enter pre-

vious code 

if code not valid

(old)

display during 

2s. and revert to measure 

display

New code registered and revert to 

measure display

if code correct

(old)

enter new code

Reminder:

 

If no key is pressed during 1 min, the instrument will revert to 

measure display. On factory exit, the access code is 0000

.

NEW

IDLE

OFF

ON

. . . .

. . . .

SIMU

GENE

ERxx

OPEN

----

• 

Deleting of the unsignificant zeros

 

                              Suppresses the display of the meaningless zeros on    

                              the right hand side.

Eg.

 : Display value 0015

                              Display  0015                                      Display      15

Eg.

 : Display value 00.15

                              Display 00.15                                      Display   0.15

• 

Self-extinguishing of the display

     

                Display permanently lit
                After 10 min, display of
In mode idle (display of 

), the following messages may appear 

briefly every 5 seconds to warn the user that the transmitter is no longer 

in a standard measurement phase. 

Displayable messages :              ,               ,              ,               or

Summary of Contents for TPI-SI 40 Series

Page 1: ...rindas Parc d activit d Arbora N 2 69510 Soucieu en Jarrest T l 04 72 31 31 30 Fax 04 72 31 31 31 Tel Intern 33 4 72 31 31 30 Fax Intern 33 4 72 31 31 31 http www ardetem com e mail info ardetem com U...

Page 2: ...gramming of a new access code with the p19 console 4 11 Functions accessible from the main menu p20 4 11 1 Simulation of the display p20 4 11 2 Simulation of the analog outputs p20 4 11 3 Deleting of...

Page 3: ...e information vertically or horizon tally according to the sense in which the converter is mounted Series TPI SI 40 440 Process inputs in standard version One bidirectionnal DC current or voltage inpu...

Page 4: ...on with graphical rear lit LCD The LCD allows visualising 4 pieces of information the value of the measure 5 mm high the unit of the displayed measure the value of the analog output or the name of the...

Page 5: ...s 0 440 and 0 2 2 k 0 8 8 k optional Potentiometer Supply for 2 wire sensor 200 850 60 260 76 500 On input mV V mA Resistive sensors and potentiometer 328 1562 200 270 from 100 to 10 k 328 518 0 1 C 0...

Page 6: ...wer supply 2 CONNECTINGS Wiring In order to keep the safety the connection must be followed strict ly especially the setting to the earth of terminal A2 of the group supply Self extinguishing case of...

Page 7: ...ogramming of the access code see p19 Display simulation see p20 Authorized by access code analog relay output analog output relay output PCOD SIMU Teleloading of a program ming file see p21 LOAD Exit...

Page 8: ...ith or p13 without saving the configuration voltage input voltage input only choice of the caliber Special linearisation only current input to t sensors resistive sensors down scale 22 00 x 22 00 mA 1...

Page 9: ...inearisation CJ E CJ T CJC YES NO ADJ To TC DISP TC L TC J TC K TCW5 TCW3 TC W TC N TC B TC E TC T TC R TC S oF oC CAL RES POT 10 0C CJC 60 0C 14 0F CJC 140 F 2 2 c Resistive sensors dINP 400 8888 FUN...

Page 10: ...ove M Downwards move decrease Menu exit access Fast version only 4 4 2 Programming of the console display inputs process resistance potentiometer position of the decimal point DISP rES dDIS 88 88 F DI...

Page 11: ...yed measure is 10 0 and 20 mA for a displayed measure of 100 0 In mode transmitter only the parameters are available For a mA input with dINP 0 00 and FINP 20 00 mA if dOU1 4 mA and F OU1 20 mA the va...

Page 12: ...Safety features p17 Safeties option 2 relays if input mV or t option analog output if return value programmed REL1 LO HI OFF OUT I state of relay 1 in case of sensor rupture LO HI OFF state of relay...

Page 13: ...YES NO REP1 value of output 1 in case of sensor rupture 00 00 rEP1 22 00 mA OFF ON 88 88 if return value OU2 I OU2 U or YES NO REP2 value of output 2 in case of sensor rupture 00 00 rEP2 22 00 mA 88 8...

Page 14: ...as well as the alarm recordings After SAVE YES la console will show WAIT during the converter re initialisation phase 4 5 1 Mode transmitter indicator According to the choice SCALE YES NO you can cho...

Page 15: ...and divide by 10 all the dis play related parameters In this example after modifying the parameter POIN from to F DIS will have to be modified from 3000 to 0300 and SP1 1 from 0300 to 0030 to obtain...

Page 16: ...00 01 04 02 85 05 04 07 47 10 00 12 54 14 96 17 17 18 98 20 00 That is to say a curve of 10 equally long segments Measure range number of segments 20mA 10 2mA length of the segment For 10 segments nb...

Page 17: ...mode Mode setpoint OFF ON setpoint Legend ON coil supplied OFF coil not supplied or MODx setpoint OFF ON SPX SPX TPI SI 41 N and AN Mode window Mode setpoint Mode window Choice of the state of the rel...

Page 18: ...entering the correct access code or by pressing simultaneously and if the access to a quick entering has been authorized on the programming of the code see p19 4 7 4 Safeties Self diagnosis The conver...

Page 19: ...7 8 9 10 400 600 1 0s 1 4s 2 0s 3 0s 5 0s 7 5s 10s 15s BRGT 1 1 1 1 4444 Setting of the digits brightness Lowest brightness Strongest brightness LdIG Inhibition of the last digit low weight In mode p...

Page 20: ...nstrument check sum of the programme no 4 9 Access code An access code adjustable from 0000 to 9999 serves to protect the converter and its setpoints from unauthorized programming and to lock the acce...

Page 21: ...analog output current output Entering of a new display value GENE GENE Value of the output if Enter the value to be injected M Press Menu to revert to measure dis play Note The instrument will carry o...

Page 22: ...display The file stored in the console can be read by validating choice READ CONS In this case the reading organigramme of the file programming is the same as for the programming of the instrument se...

Page 23: ...or 500 C the display will be 0500 2 To correct this display the display down scale must be shifted by 3 points to obtain 000 0 and the display full scale by 2 points to obtain 0500 0 Press to validate...

Page 24: ...ed simply press and the converter will revert to mode measure taking the new values into account If no key is pressed after 60 s the converter will revert to the measure display without modification o...

Page 25: ...enu Input type Voltage input Current input Temperature input Potentiometer input Resistance input Potentiometer and resistance inputs Input down scale Input full scale Choice of the processing functio...

Page 26: ...esis in display points Time delay on relay X OFF MODx _ _I I_ _ AL x REL x ON REL SPx 1 SPx 2 HYSx TIM x SPx _I I_ I_ _I SCAL POIN dDIS FDIS Yxx CUT 10 FILT UNIT T MES RES 0 10 Display parameters Subm...

Page 27: ...upture safety Validation or not of the sensor rupture Sensor rupture inactive Sensor rupt active State of relay X in case of sensor rupture No sensor rupture associated with the relay Relay de activat...

Page 28: ...ress 200 202 204 208 210 212 290 291 Value of the analog output n 1 on TPI SI 440 441 in A mA output Minimum value of the displayed value Maximum value of the displayed value Direct measure Value of t...

Page 29: ...3 Nbr of MSB words LSB 2 bytes 2 bytes Slave number 1 byte 1 byte 1 byte Function 3 or 4 2 bytes CRC 16 Number of read bytes Response pattern 1st word MSB value LSB 2 bytes 2nd word MSB value LSB 2 by...

Page 30: ...byte 1 x 256 byte 2 0 x 2563 0 x 2562 19 x 256 136 5000 Reading of the address 120 decimal point 2 displayed measure 50 00 9 4 CRC 16 calculation algorythm FFFF CRC CRC BYTE CRC END yes no n 0 n n 1 y...

Page 31: ...s The directive EMC 2014 30 UE generic standards IEC 61000 6 2 Immunity for industrial environments IEC 61000 6 4 Rejections for industrial environments The directive ATEX 2014 34 UE standards EN 6007...

Page 32: ...p31...

Reviews: