If the freezer has not been programmed on
set point in more than one hour, the display
shows “not ready”. Otherwise the
calibration menu is shown.
The technician can set the code, which
gives access to the service menus.
The technician can programme the set
point of the freezer.
The programming of the temperature
hysteresis. Compressor starts at TSet +
TDiff and is turned off at TSet.
The programming of the lowest alarm limit.
TA1Lo + TSet= Alarm
The programming of the highest alarm
limit. TA1Hi + TSet = Alarm
The delay of the abovementioned alarms
starts.
The technician has to programme the
maximum allowed room temperature
before the room temperature alarm is
activated.
The programming of the shortest time
which is allowed between every
compressor start. Depends on the type of
the compressor.
Allows you to change the code.
Language menu for the menu system. The
text which describes the language is
always on the same language. For
example: (English, Dansk, Deutsch…)
Can be bought as an option.
Back to menu.
-