Explanation of Alarms:
AL10 - T1 OPEN
: Temperature sensor to freezer interrupted
AL11 - T1 SHORT
: Temperature sensor to freezer short circuited
AL12 - T2 OPEN
: Temperature sensor to room interrupted
AL30 - MAINS OFF
: No mails supply to the freezer
AL31 - DOOR
: Dor was open for more than 5 minutes
AL32 - ROOMTEMP
: Room temperature too high
AL33 - T LOW
: Freezer temperature too low for more time than alarm delay
AL34 - T HIGH
: Freezer temperature too low for more time than alarm delay
AL40 - T.SENSOR
: User test fault at room temperature sensor
AL41 - D.SENSOR
: User test fault at door sensor
AL42 - NO SOUND
: User test fault at buzzer
The Cooling System
The vaporiser is built into the shelves of the inner container. They will be approx. -
85
°C, so there is a risk of
frostbite if they are touched. Always use thermal gloves when touching the contents of the freezer.
The condenser is placed on top of the cabinet behind the control panel.
Thermostat with digital display: once the freezer is cooled down, the set temperature and the temperature
shown will be the same.
CO
2
Back Up
2
4 x port holes are prepared for CO back up, and an injection tube (see separate enclosure) can be inserted
2
from the top of the unit (remove kit). Make sure that pressurized CO escape from the gaskets or pressure
valve.
Service with
for - °C freezers
Note that the
EP88
gas can only be filled from a bottle containing a measured amount. Normal procedures
used in conjunction with other refrigerants cannot be used.
Further note that the freezing capability is very dependent upon filling the entire contents of the bottle into
the system. Freezing capability is also very dependent upon the temperature of the
EP88
and whether it
has been still for a long period of time. The temperature of the service bottle must be between 8 and 14
degrees and the contents mixed by shaking the bottle 10-15 times.
In order to carry out the service procedure correctly, it is necessary to use a capillary pipe with mounting
screws between the bottle and the filling station instead of a normal filling hose. The service technician may
do this by using the mounting screws of a normal filling hose onto a capillary pipe.