5.7
Replacement of Stop-Flow mechanism
The Stop-Flow mechanism opens and closes the tubing automatically to prevent free flow
if the door is opened by accident. Its function is self-checked every time an infusion is
started.
Tools & recommended spare parts:
- Antistatic work place
- 2.5 mm Hex driver
- 5.5 mm socket wrench
- Philips screw driver No. 1
- Stop Flow kit ( Robson adapter 76094, Internal adapter 76096)
Procedure:
1.
Dismantle main PCB
2.
Dismantle pump block
3.
Loosen screws as shown on picture
4.
Carefully remove front part of Stop flow Mechanism
Pull bottom of adaptor out
to release bottom edge
Pull adaptor down to
release from Actuator
Adaptor is now safe to
remove