SH-7000-Pre-E-04/16
20
5.2
Dismantling the Door
Tools & recommended spare parts:
- Antistatic Work place
- 2,5 mm Hex driver
- 1 mm Philips screw driver
- Door Complete (76390 PVC/EN, 76370 Silicone/EN))
Procedure:
1.
Remove the main PCB.
2.
Unplug the display ribbon cable from main PCB.
3.
Loosen the two screws indicated and carefully remove door from metal rod.
4.
Grip rubber seal and remove from housing. (Push out to the front), now the door is
free. Pull the door cable out of housing.
5.
Remove the silver stickers; do not remove the clear sticker of the air detector plate.