PL7004 Kit
26
Rev F
Command To Probe Kit
Response From Probe Kit
Set Response Termination
Character(s)
TERMn<CR>
Query Response
Termination Character(s)
TERM?<CR>
Where:
n
=
0
to
3
0
=
<LF><CR> factory default
(for compatibility with legacy control
software)
1
=
<CR><LF>
2
=
<LF>
3
=
<CR>
Note 1: After this command is sent, a query response is automatically
generated reflecting the present termination setting.
Note 2: Setting is retained when power is cycled.
Note 3: Selection 0 must be used for compatibility with FM7004A and
VM7000.
Response Structure:
TERMn<termination>
(see above for variable definition)
termination
= Depends on the selected response termination
characters (1 or 2 characters)
3.7.1.2
Error Code Responses
Error code responses are transmitted from the probe kit due to errors detected during operation upon receipt of
a serial input. The general format is similar to other responses. When an error has been recognized and an
error code has been generated, the probe kit will generally check and then clear its input buffer until a new
start character is received. Specific responses are as listed below. Recommended actions are shown in
brackets [ ]. <termination> depends on the selected response termination characters (1 or 2 characters).
Format:
Ex
where
x = a….z
Code
Command
Detail
Ea<termination> Framing error
Indicates that corrupted data has been recognized by the probe kit,
and that no further processing is to be done on this data string.
Eb<termination> Input buffer error
Overflow, command has been lost due to probe failure to process
some input character in time and the input character has been lost.
[Either there is a problem with the probe kit, or data is coming too
fast and this advises that host should send characters with more
time between characters.]
Ec<termination> Communication
format error
Start character received, but no stop character <CR> received
within approximately 5 seconds of a start character, and probe
input buffer has been cleared. The long time is to allow for a
command to be entered by an operator using a keyboard. [Host
should retransmit the previous command and insure that future
Summary of Contents for FI7000
Page 2: ......
Page 4: ......
Page 5: ......
Page 6: ......
Page 20: ......
Page 24: ......
Page 28: ...PL7004 Kit 12 Rev F...
Page 46: ...PL7004 Kit 30 Rev F...
Page 50: ...PL7004 Kit 34 Rev F...
Page 51: ...Rev F 35 Appendix A Dimensional Drawings Figure A 1 Model PL7004...
Page 52: ...PL7004 Kit 36 Rev F...
Page 54: ......