<COND/TDS>

Make sure the sensor is 3-in-1 electrode and switch to 
pH mode.

Dip the electrode in

to the buffer solution 1413 ìS/cm 

(706 PPM). Stir gently and wait until the reading is 
stable. Press and hold “

CAL

” for 5 sec. to enter 

calibration mode. The display will appear 

CAL

 and 

flashing 1413 ìS/cm (706 PPM). When the display stops 
flashing calibra

tion ends, and meter will return to 

measurement mode.

Note:

Calibration error indicator icon will appear, and “Err” if 
calibration fails.

Measurement:

 

<pH>

Press “

MODE

” to choose pH mode.

After calibration, rinse the 3-in-1 electrode with clean 
water and wipe it dry. Dip the electrode into the sample 
solution which is going to be measured. Stir gently and 
wait until a stable reading can be obtained.