Accessories :
Meter with 3-in-1 electrodePouchInstruction manualLR44 battery x 3 (installed in the meter)
Measure Readings
Temperature
Display Description:PH
Battery Status
Unit
Conductivity
Parameter
TDS
Page 1: ...Multi Parameter Handheld Meter pH Conductivity TDS Salt Temp User s Guide AM P AL...
Page 2: ...unsuitable operation and maintenance Our Handheld meter employs leading edge technology with integrated microprocessor which is suitable for measurement in water solutions for institutes industrial l...
Page 3: ...ssories Meter with 3 in 1 electrode Pouch Instruction manual LR44 battery x 3 installed in the meter Measure Readings Temperature Display Description PH Battery Status Unit Conductivity Parameter Unit...
Page 4: ...t Press to change Temperature unit Long Press for Calibration To Change Mode between pH EC TDS Preparation Remove the protection cap from the electrode Rinse the electrode with clean water and wipe it...
Page 5: ...s and meter will return to measurement mode Rinse the electrode with clean water and wipe it dry Dip the electrode into the buffer solution pH 4 01 and follow calibration as previous steps Rinse the e...
Page 6: ...CAL and flashing 1413 S cm 706 PPM When the display stops flashing calibration ends and meter will return to measurement mode Note Calibration error indicator icon will appear and Err if calibration...
Page 7: ...tery Change 1 If battery is low meter will flash low indication 2 Open the battery case take out the old battery and replace with new one 3Nos x LR44 cells 3 Please mind the polarity of the batteries...