www.aquantis.org
Microfreeze User’s Guide
19
7.3.4.2
Dashboard
Protection level: freely accessible.
F
IGURE
10:
D
ASHBOARD WEBPAGE
The home-page of the MICROFREEZE sensor consist of a freely accessible dashboard (Figure 10) on which the
following parameters of the sensor are published:
1.
PRODUCT SELECTION: shows the active product together with its product label as specified in the
product database. Pressing this button activates a drop-down menu in which a product from the product
database can be selected.
2.
OPERATION MODE: shows the mode of operation of the sensor. In
static
mode of operation, the sensor
is continuously publishing new measurement values. This mode is intended to be used for at-line
measurements where the product under test is statically placed underneath the
sensor’s
aperture. In
dynamic
mode of operation, the sensor is analysing each interaction of passing products. Only when
detected products are accepted for measurement (based on criteria specified in the product database),
the new measurement is published. In this mode, the product averaging is active, and you can select
the averaging number of the accepted products.