“Calibration needed”
No product (thickness < the range specified in product database)
Low signal (signal too small for no product
–
< 80% of current calibration level)
solution:
run sensor and thickness calibration wizards
“
No Product or product to thin
”
Measured product thickness < minimum product thickness specified in the
product database
causes:
- Wrong product selection or database parameters
- Product is not placed correctly under the sensor
- Thickness sensor needs calibration
–
execute the thickness calibration
wizard
“Product
to thick
”
Measured product thickness > maximum product thickness specified in the product
database
causes:
- Wrong product selection or database parameters
- Multiple products are stacked on top of each other
- Thickness sensor needs calibration
–
execute the thickness calibration wizard