4. Maintenance 
 

Flushing: 

In order to flush the membrane, only the flushing valve has to be opened. Then, the 

water flows with high speed through the narrow channels in the membrane and takes most of the 
residuals with it. This flushing also cleans the flushing valve itself.  
 
After  15  -  20  minutes,  the  valve  must  be  closed  again  -  tear  only  handtight.  If  you  tighten  the 
valve  too  hard,  the  valve  and  membrane  will be destroyed. A minimum water flow (approx. 12 - 
15/ 15 - 18 l/h) has to be guaranteed. 
 
-

 Pre-filter:

 If the tap water is not extremely polluted, a yearly replacement of both pre-filters will 

be  sufficient.  If  the  water  is  higher  polluted,  the  filters  have  to  be  changed  more  often  -  as 
required. If the pre-filter is blocked, the water production will slow down. 
 
-

 Membrane:

 The membrane has a lifetime of several years. Afterwards, it has to be replaced. 

 

Flushing  valve:

  The  flushing  valve  is  a  part  with  limited  lifetime.  In  case  of  a  decrease  of  

adjusted water flow, the valve is encrusted with calcium and can be cleaned using acid (e.g. acetic 
acid). If the water flow increases, the valve has to be replaced. 
 

5. Failures  
 
- Waste water flow rate is too high:

  

Try to close the flushing valve. If this is not possible, the flushing valve must be replaced and check 
the membrane power.   
  

 

- Waste water flow rate is too low:

 

Check if the pre-filter is blocked. The water pressure at the pressure gauge behind the pre-filters 
must not be much lower (at the most 0.5 bar) than the water supply pressure. Maybe change the 
pre-filter.  

 
- Pure water flow rate is too low:

 

Check the waste water flow rate. If it is to low too, change the pre-filter. 

 
- Tap water pressure is too low: 

If  you  are  connected  to  a  municipal  water  supply,  you  are  not  able  to  have  an  influence  on  the 
water pressure. For the reverse osmosis unit, a minimum water pressure of 3 bar is required. 

 

- Module is blocked or used up: 

The membrane has to be replaced. 

As spare parts, all standard „18“ membranes with a daily capacity of 75 - 90 / 120 - 150 l can be 
used. Only TFC-membranes made of Polyamide/Polysulfone should be used. We recommend to use 
original AB Aqua Medic TFC-membranes only. If you want to use a membrane with a different 
capacity, you have to change also the flushing valve. 

 

- Rejection rate is too low: 

Check the water flow rates (waste water, pure water). 
If  the  water  flow  is  ok,  flush  the  unit  for  approx.  60  min.  If  the  rejection  rate  increases,  go  on 
flushing until the value is acceptable. If the rate does not improve, change membrane. 

 
6. Important Notes 
 
-  Connection: 

Connect  the  reverse  osmosis  unit  to  cold  water  supply  only.  Higher  temperatures 

increase  pure  water  production  but  at  temperatures  more  than  40  °C,  the  membrane  will  be 
destroyed. 
 

Summary of Contents for Premium 90/150

Page 1: ... in der Lage die anorganischen Inhaltsstoffe und andere Schadstoffe Ihres Leitungswassers ohne Zudosierung von Chemikalien um 95 98 Nitrat und Silikat ca 70 zu vermindern Lesen Sie die Bedienungsanleitung sorgfältig Dies vermeidet Fehlbedienungen und Reklamationen GmbH Gewerbepark 24 49143 Bissendorf Germany __________________________________________________________________________________________...

Page 2: ...tungswasserzulauf in den Vorfilter 1 2 Konzentratablauf Abwasser am Spülventil 3 3 Reinwasser Permeat 2 4 Wasserhahnanschluss Den Wasserhahnanschluss an einen 3 4 Wasserhahn Kaltwasser anschließen ggfs Adapterstück zwischenschalten nicht im Lieferumfang enthalten 3 Inbetriebnahme Wenn alle Anschlüsse fest sitzen kann das Wasser aufgedreht werden Der minimale Wasserdruck für den Betrieb der Anlage ...

Page 3: ...lventil 4 Schraubkappe Umkehrosmosemodul 5 Umkehrosmosemodul mit Membrane 6 Vorfiltergehäuse für Feinfilter 7 Vorfiltergehäuse für AK Filter 8 O Ringe 9 Flachdichtung 10 Feinfiltereinsatz 11 Aktivkohlefiltereinsatz 12 Halteklammer Modulgehäuse 13 Befestigungswinkel ...

Page 4: ... Ggfs Vorfilter wechseln Es fließt zu wenig Reinwasser Prüfen ob auch zu wenig Abwasser fließt und ggfs Vorfilter wechseln s o Wasserdruck zu niedrig Hier können Sie als normaler Wasserverbraucher wenig machen wenn der niedrige Wasserdruck schon vom Wasserwerk geliefert wird nachfragen Für die Anlage sollte der Wasserdruck mind 3 bar betragen Modul verstopft oder verbraucht Die Membrane muss erneu...

Page 5: ...ienachweis gilt der Original Kaufbeleg Während dieser Zeit werden wir das Produkt kostenlos durch Einbau neuer oder erneuerter Teile instand setzen ausgenommen Frachtkosten Im Fall dass während oder nach Ablauf der Garantiezeit Probleme mit Ihrem Gerät auftreten wenden Sie sich bitte an Ihren Fachhändler Diese Garantie gilt nur für den Erstkäufer Sie deckt nur Material und Verarbeitungsfehler die ...

Page 6: ...and silicate approx 70 without adding any chemicals In the aquarium you prevent overfertilization of the water from the tap The growth rate of diatoms and other microalgaes is reduced Residues of pesticides and other toxic substances do not get into the aquarium Please read the manual carefully This prevents failures and complaints GmbH Gewerbepark 24 49143 Bissendorf Germany _____________________...

Page 7: ...et into the sediment pre filter 1 2 Concentrate outlet waste water at flushing valve 3 3 Pure water permeat 2 4 Tap water connection Fix the water connection piece to a 3 4 water tap cold water If it does not fit use a reduction piece not included in shipment 3 Starting When all connections are tight you can open the water tap The minimum water pressure for the unit is 3 bar The nominal capacity i...

Page 8: ...p of module housing 5 Reverse osmosis module incl membrane 6 Pre filter housing for fine filter 7 Pre filter housing for activated carbon filter 8 O rings 9 Flat gasket 10 Fine filter element 11 Activated carbon filter element 12 Holding clip for module s housing 13 Mounting bracket ...

Page 9: ...and check the membrane power Waste water flow rate is too low Check if the pre filter is blocked The water pressure at the pressure gauge behind the pre filters must not be much lower at the most 0 5 bar than the water supply pressure Maybe change the pre filter Pure water flow rate is too low Check the waste water flow rate If it is to low too change the pre filter Tap water pressure is too low I...

Page 10: ...ble products Proof of Purchase is required by presentation of an original invoice or receipt indicating the dealer s name the model number and date of purchase or a Guarantee Card if appropriate This warranty may not apply if any model or production number has been altered deleted or removed unauthorised persons or organisations have executed repairs modifications or alterations or damage is cause...

Page 11: ...rofessionnels Cet appareil vous permet de retirer en moyenne de l ordre de 95 à 98 les substances inorganiques et autres impuretés nitrates et silicates 70 présentes dans votre eau de conduite sans avoir recours à des produits chimiques Veuillez lire soigneusement les instructions Ceci empêche les erreurs de manipulation et les réclamations GmbH Gewerbepark 24 49143 Bissendorf Germany ____________...

Page 12: ... concentrat au niveau de la soupape de rinçage 3 3 Perméat eau osmosée 2 4 Raccordement robinet eau de conduite Connecter le raccordement du robinet d eau de conduite avec un robinet type 3 4 ou intercaler un réducteur non compris dans l emballage 3 Mise en route Lorsque tous les raccordements sont effectués il est possible d ouvrir l arrivée d eau La pression minimale pour le fonctionnement de l ...

Page 13: ...çage 4 Couvercle à vis module d osmose 5 Module d osmose avec membrane 6 Boîtier du préfiltre pour filtre fin 7 Boîtier préfiltre pour filtre charbon actif 8 Joint 9 Joint plat 10 Filtre 11 Filtre à charbon actif 12 Pince de fixation boîtier du module 13 Equerre de fixation ...

Page 14: ...eté élevée l appareil peut très rapidement se calcifier et il faut alors le retourner à l usine Il n y a pas assez de concentrat Vérifier si le préfiltre est colmaté La pression du manomètre situé derrière le préfiltre ne doit pas être inférieure 0 5 bar à celle de la pression de l eau de conduite voir système de coupure Le cas échéant changer le préfiltre Il n y a pas assez de perméat Vérifier si...

Page 15: ...eil doit être stocké à l abri du gel Le gel détruit la membrane 7 Garantie AB Aqua Medic GmbH assure une garantie de 12 mois à partir de la date de l achat sur tous les défauts de matériaux et d assemblage de l appareil Elle ne couvre pas les pièces d usure comme le tube UV C ou la gaine de quartz Le ticket de caisse original sert de preuve d achat Durant cette période l appareil est remis gratuit...

Reviews: