Chapter 5
Grounding and Cabling the IONizer
Overview
This chapter provides grounding and cabling information for IONizers in the 4200 Series.
Grounding
All IONizers must be properly grounded before making power and signal connections. Apprion
recommends a UL Listed #10 AWG wire that is suitable for this purpose, and a ring tongue terminal,
Panduit P/N P10-8R or equivalent. The terminal is to be crimped to the wire using the correct crimping
tool as recommended by the terminal manufacturer. The torque rating on the nut is 8 to 10 inch
pounds. The wire should be kept as short as possible while using grounding practices that are compliant
with local codes and practices.
Note:
Ensure that the connection to a proper earth ground is made by certified and authorized
personnel. The ground must conform to all applicable codes and regulations. The materials required to
connect to a proper earth ground are defined by local conditions and must be procured locally to ensure
that the correct safety environment is achieved.
Grounding Procedures (IONizer 4200, 4220)
The grounding stud is located on the bottom of the IONizer at the location shown:
Attach the earth ground wire (not supplied) to the ring terminal attached to the IONizer’s grounding
stud. Ensure that the ring terminal is seated against the IONizer’s metal case.