Chirascan Couette Cell LD Accessory User Manual
3.2
Installing the Couette Cell LD Accessory
3.2.1 Installing the accessory body
Before the accessory is installed, the standard bulkhead blanking plate should be removed from the Chirascan
sample chamber and replaced with the plate fitted with two power connectors (Figure
: note that in the
figure the blanking plate also contains a fitting for the Chirascan Titration accessory). The plate is held in
position from inside the sample chamber by four caphead screws.
Figure 3.2: The blanking plate mounted on the Chirascan sample chamber
The accessory stands on a mounting plate which locates on a support plate within the Chirascan sample
chamber. Place the mounting plate of the accessory over the support plate in the sample chamber. The plate
seats down with the holes at the front left and rear right locating over the corresponding pegs on the support
plate. When the plate is seated down, secure it in place by tightening the two captive thumbscrews positioned
at the right front and left rear of the mounting plate.
The detector can now be re-installed. The snout of the detector slides into the cowling which rests against the
accessory body
. The cowling has a gas purge inlet, which connects to the ‘Y’ connector at the rear of the unit,
using the additional steel tubing provided. The extendable threaded spacer fits between the light inlet port and
the accessory. The spacer has O-rings at each end: the narrower end goes towards the accessory assembly.
Connect the Legris type connectors for the circulating water by pushing home the male on the accessory to
the female on the sample chamber.
Ensuring that the accessory electronics box is not powered up, connect the two circular electrical connectors.
Note that the smaller, 3-pin connector serves the LED and the larger, 8-pin connector serves the motor.
When the accessory is fully installed without the Couette cell, the interior of the spectrometer sample chamber
should appear as in Figure
Blanking plate