Document 4207Q258 V3.00 January 2018
Page 16 of 30
Chirascan Couette Cell LD Accessory User Manual
Figure 2.7: the LDConf.conf file
If the measured absolute value of the LD at 250 nm is not 0.03006 ± 0.0003 (i.e. within 1% of the accepted
value), the calibration factor should be adjusted. The new calibration factor is calculated as:
New calibration factor = present calibration factor x 0.03006 / measured LD value.
By convention the parallel direction for the Couette cell is taken as being the flow direction, i.e. the horizontal
direction of the alignment jig. If the LD calibration factor gives the correct absolute value, but the incorrect sign,
change the sign of the calibration factor.
Input the new factor and save the file.
2.3
References
1,
B. Norden and S. Seth, Appl. Spectroscopy, 39 (1985) 647-655.