Basic Operation
Figure 4-8
V-ON Setup Page
Step 2
: Enter the V-ON Value by using numeric keypad or knob. Enter “0” can disable
V-ON function.
Step 3
: Press
Enter
to confirm input and the V-OFF Setup Page displayed.
Figure 4-9
V-OFF Setup Page
Step 4:
Enter the V-OFF Value by using numeric keypad or knob. Enter “0” can
disable V-OFF function.
Step 5:
The VLMT indicator will be showed on VFD once Von or/and Voff value
was/were set.
4.7
Load ON Timer Function
When the “Load ON Timer” enabled and timer value reach the setup time, AT851 will
turn off the input automatically.
“Load ON Timer” Setup Procedure:
Press
Shift
+
Timer
to enter “Load ON Timer” setup Page.
Figure 4-10
Load ON Timer Setup
Page
Enter the timer Value by using numeric keypad or knob.
Enter “0” can disable “Load ON Timer” function.
The unit is “second” and maximum time is 60000s
。
4.8
File Operation (Saving/Loading/Erasing Settings)
You can save these following values into internal non-volatile memory:
I-set value and I-max value
V-set value and V-max value
P-set value and P-max value
R-set value and R-max value
Transient setting value
AT851 provides 20 files (file No. from 0 to 19). You must save setting load parameters
into files manually.
If the Menu-RECALL Item set to “ON”, the saving setting file will be load into system