![Applent Instruments AT520 User Manual Download Page 22](http://html.mh-extra.com/html/applent-instruments/at520/at520_user-manual_2954882022.webp)
AT520/AT520SE/AT520L/AT520M User’s Guide
22
Step3: Press Enter key to exit with setting value saved.
Press ESC key to exit without being saved.
5.4
How
the
comparator
work
When comparator is ON,
Resistance:
Limit Lower < Rx/< Limit Upper IN
Rx
Limit Lower
LO
Rx
Limit Upper
HI
Table 5-1
Comparator Output (Resistance/Voltage measurement mode)
Table 5-2
Comparator setting example:
Accordance with the above settings to test a battery, the measured results of resistance:
100 m
, voltage 1.40 V, then sorting the results of its output is "NG IN".
Accordance with the above settings to test a battery, the measured results of resistance:
100 m
, voltage 1.51 V, then sorting the results of its output is "GD IN".
Accordance with the above settings to test a battery, the measured results of resistance:
150 m
, voltage 1.51 V, then sorting the results of its output is "NG Hi".
Accordance with the above settings to test a battery, the measured results of resistance: 60
m
, voltage 1.50 V, then sorting the results of its output is "NG Lo".
Resistance
Voltage
Hi IN Lo
Hi
NG NG NG
IN
NG GD NG
Lo
NG NG NG
Resistance Nominal Value
Upper Limit
Lower Limit
100 m
120
m
80
m
Voltage Nominal Value
Upper Limit
Lower Limit
1.5 V
1.52 V
1.48 V