AT520/AT520SE/AT520L/AT520M User’s Guide
18
Press Shift + Range to select the voltage range (AT520/AT520SE/AT520L).
No. Range Maximum
indication
Resolution
1 6V
6.0000V 100
V
2 60V
60.000V
1mV
AT520M
No. Range Maximum
indication
Resolution
1 10V
10.000V
1mV
2 100V
100.00V
10mV
Note
1. When auto-range is selected, both resistance and voltage ranges are set to auto-range
mode.
2. If the range is not determined in auto-range mode, perform a zero-adjust.
4.3
Setting
the
Sampling
Rate
The RATE operation sets the integration time of the A/D converter, the period of time the
input signal is measured (also known as aperture). The integration time affects the usable
digits and the amount of reading noise.
The RATE items are explained as follows, you can press Rate key to choose.
Fast
: 20 readings/s. Use FAST if speed is of primary importance, at the expense of
increased reading noise and fewer usable digits.
Medium
: 6 readings/s. Use Medium when a compromise between noise performance and
speed is acceptable.
Slow:
1 readings/s. SLOW provides better noise performance at the expense of speed.
4.4
Short
Clear
Zero
Correction
The SHORT correction function reduces the measurement error due to the residual
impedance existing between the UNKNOWN terminals and the DUT.
Step1: Under measurement mode, connect the leads as shown below:
Right connection:
Wrong connection:
Step2: Press Clear Key when the value observed in the VFD stabilizes.
Step3: After Clear Zero Correction, AT520x returns to measurement mode.
4.5
Choose
Measure
Parameter
(
AT520
)
Step1: Under measurement mode, Press Shift Key. “Shift” appears lit on the VFD.
Step2: Press Rate (I-Mode) key.
Step3: To switch measure parameter as below:
R-
X-