Symbols as marked on the meter and Instruction manual

 

 

Risk of electric shock 

 

See instruction manual 

 

DC measurement 

 

Equipment protected by double or reinforced insulation 

<

 

Battery 

 

Earth 

2

 

AC measurement 

 

Conforms to EU directives 

 

Do not discard this product or throw away 

E

 

Application around and removal from hazardous live conductors is permitted 

Summary of Contents for 30R

Page 1: ...AC DC CLAMP MULTIMETER INSTRUCTION MANUAL APPA 30R ...

Page 2: ...e barrier when in normal use Do not use a flexible current sensor if the inner contrasting color of the insulation of the flexible cord is visible If the equipment is used in a manner not specified by the manufacturer the protection provided by the equipment may be impaired CAT Ⅱ Is for measurement performed on circuits directly connected to the low voltage installation CAT Ⅲ Is for measurement pe...

Page 3: ...shock See instruction manual DC measurement Equipment protected by double or reinforced insulation Battery Earth 2 AC measurement Conforms to EU directives Do not discard this product or throw away E Application around and removal from hazardous live conductors is permitted ...

Page 4: ...ing and Inspection Upon removing your new Digital Clamp Multimeter from its packing you should have the following items 1 Digital Clamp Multimeter 2 Test lead set one black one red 3 Carrying case 4 Instruction manual 5 Battery ...

Page 5: ... This slide switch is used to select V V A A Ω function 5 P MAX Switch This switch has two modes one is data hold the other is maximum hold Sliding the function switch to power on this switch will work in data hold mode Pressing the P MAX Switch then slide the function switch from power off to power on this switch will work in maximum hold mode Data hold mode This mode is used to hold measured val...

Page 6: ...easure resistor or continuity in Ω function Auto power off disable mode Press this switch until slide the function switch power on for 1 second it will cause the auto power off function disable 8 Trigger Press the lever to open the transformer jaws When the pressure on the lever is released the jaws will close again 9 Hand Guard Designed to protect user for safety 10 Transformer Jaws Designed to p...

Page 7: ...6 Figure 1 ...

Page 8: ...Over range Indication OL indicated Low Battery Indication is displayed when the battery voltage drops below operating voltage Measuring Rate 2 times second normal Position Error 1 of reading Type of Sensing Hall effect sensing for AC and DC current Shook Proof 4 feet drops Power Requirement Alkaline AAA size 1 5V x 2 Battery Life Alkaline 100 hours Maximum Jaw Opening 25mm ...

Page 9: ...05 grams including battery Accessories Test leads battery manual and carrying case 2 2 Environmental Conditions Indoor use Maximum Altitude 2000 Meter Installation Category IEC 61010 CAT II 600V CAT III 300V Pollution Degree 2 Operating Temperature 0 C to 30 C 80 RH 30 C to 40 C 75 RH 40 C to 50 C 45 RH Storage Temperature 20 C to 60 C ...

Page 10: ...than 80 R H 1 AC Voltage Auto ranging Range Resolution Accuracy Over voltage protection 400 0mV 100μV 2 0 rdg 5dgt 50Hz 60Hz 600V rms 4 000V 1mV 1 5 rdg 5dgt 40Hz 300Hz 40 00V 10mV 400 0V 100mV 600V 1V 1 5 rdg 5dgt 40Hz 500Hz Input Impedance 10MΩ less than 100pF Less than 30 digital rolling ...

Page 11: ...calibrated to the rms value of a sine wave input Accuracies are given for sine wave at full scale For distorted signals add the following Crest Factor corrections For Crest Factor of 1 4 to 2 0 add 1 0 to accuracy For Crest Factor of 2 0 to 2 5 add 2 5 to accuracy For Crest Factor of 2 5 to 3 0 add 4 0 to accuracy ...

Page 12: ...11 2 DC Voltage Auto ranging Range Resolution Accuracy Over voltage protection 400 0mV 100μV 0 5 rdg 5 dgt 600V rms 4 000V 1mV 40 00V 10mV 400 0V 100mV 600V 1V 0 5 rdg 2 dgt Input Impedance 10MΩ ...

Page 13: ...tance Auto ranging Range Resolution Accuracy Overload Protection 400 0Ω 100mΩ 1 2 rdg 6 dgt 1 600V rms 4 000KΩ 1Ω 0 9 rdg 3 dgt 2 40 00KΩ 10Ω 400 0KΩ 100Ω 4 000MΩ 1KΩ 40 00MΩ 10KΩ 2 5 rdg 5dgt 1 3 1 2 rdg 3 dgt 2 ...

Page 14: ...nse time is approximate 20 seconds Put a low resistor in the input terminal before slide the sliding function switch to resistor and continuity function maybe cause a buzzer sound 4 Continuity Built in buzzer sound when measured resistance is less than 30Ω and sound off when measured resistance is more than 300Ω Between 30Ω to 300Ω the buzzer maybe sound or off either ...

Page 15: ...cy Over voltage protection 0 40 00A 10mA 400A rms 40 0A 200 0A 100mA 200 0A 300 0A 100mA 2 0 rdg 2 dgt 1 0 rdg 2 dgt For DCA ACA 1 Temperature Coefficient 0 2 x Spec Acc y C 20 C or 26 C 2 Operating Temperature 0 C to 30 C 80 RH 30 C to 40 C 75 RH ...

Page 16: ...Overload Protection 0 4 00A 10mA 1 0 rdg 5 dgt 50Hz 60Hz 400A r m s 4 00A 40 00A 10mA 1 0 rdg 3 dgt 40 0A 200 0A 100mA 200 0A 300 0A 100mA 3 0 rdg 3 dgt 0 4 00A 10mA 2 0 rdg 7 dgt 40Hz 1KHz 4 00A 40 00A 40 0A 200 0A 100mA 200 0A 300 0A 100mA 5 0 rdg 5 dgt 2 0 rdg 5 dgt ...

Page 17: ...ld 7 1 In maximum hold function the accuracy is changed as following Original Accuracy 10 digitals change steps of range For example At First the maximum hold reading on display is 100 0mV on 400 0mV range f a voltage vibration changes the maximum hold reading to 120 0V The change steps of range are 3 steps 400 0mV to 4 000V to 40 00V to 400 0V so the accuracy is needed to add 3 steps x 10 digital...

Page 18: ...agnetic interference the display may be unstable or indicate incorrect measurement values 2 Make sure that the battery is properly connected 3 The instrument should only be operated between 0 C 50 C and at less than 80 R H except current function is operated between 0 C 40 C 4 Do not use or store this instrument in a high temperature or high humidity environment and do not store the unit in direct...

Page 19: ...positioned around only one conductors of a circuit If the clamp is placed around two or more current carrying conductors the meter reading will be FALSE 4 For DC measurement the reading is positive value when the current flows from the upper side to the lower side of the instrument as Fig 2 5 Using the zero switch to zero the reading Due to the high sensitivity of the clamp meter must zero in the ...

Page 20: ...19 Figure 2 ...

Page 21: ...20 Figure 3 ...

Page 22: ...hing the AC DC switch to select AC mode or DC mode 3 4 Resistance Measurement 1 Set the slide switch at Ω position 2 Connect the black test lead to the COM terminal and red lead to the V Ω terminal 3 Verify that the power to the circuit under test is off Connect test leads to the circuit to make the measurement 4 Pushing the Ω switch to select Resistance mode or Continuity mode 5 At mode Built in ...

Page 23: ...vents 4 2 Battery Installation or Replacement The meter is powered by two 1 5V alkaline battery Refer to Fig 4 and use the following procedure to replace the battery 1 Disconnect the test leads and turn the meter off Remove the test leads from the front terminals 2 Remove the battery cover of case bottom from the instrument by removing the screw and then lifting off battery cover 3 Lift the batter...

Page 24: ...23 Battery Cover Screw Battery Figure 4 ...

Page 25: ...1 Pao Zong Rd Shin Tien Taipai 23115 Taiwan R O C P O Box 12 24 Shin Tien Taiwan Tel 886 2 9178820 Fax 886 2 9170848 E MAIL info appatech com http www appatech com Printed In Taiwan Copright 2013 APPA Tech Corp All rights reserved ...