5.4.8
DATA LOG SETUP
Automatic logging of test data results is supported by the DLTM4. The Data Log Setup display in Figure 16 permits the
user to configure this automatic feature.
F1
F2
F3
F4
F5
Figure 16: Data Log Setup Display
The current date and time can be entered by the user if needed. There are five measurements that are displayed on the Data
Log Setup in the Measurement 1-5 fields are reported following the date and time. These fields can be any of the fields
from the Status Display (in Table 3).
Data Log records are transmitted through the remote programming port
Table 9: Selectable Definitions of the actions field on the Data Log Setup Display
Field
Definition
Log Type
This field is set by the user to select the data logging mode. Three selections are offered: None,
Interval and On Change. Selecting None disables the Data Log process. The Interval mode
causes a report to be generated at the expiration of the number of seconds specified in the LOG
TIME field. Selecting the On Change mode causes a report to be generated each time a
measurement changes from its previous value. The On Change mode should not be used if
measurements are expected to change at a rapid rate.
Date
Operator entry field for the current date in mm/dd/yy format
Time
Operator entry field for the current time in hh/mm/ss format
Interval
Interval in seconds between data log records if the DATA LOG field is set to Interval
Measurement
n
This field allows selection of any measurement from the Status display for logging
Function Keys
F1
Go directly to Measurements display
F2
Go directly to the Output Settings display
F3
Go directly to the Input Settings display
F4
Go directly to the Presets display