Apogee Instruments TS-100 Owner'S Manual Download Page 14

MAINTENANCE AND RECALIBRATION 

At full duty, the fan is designed to last 50,000 hours (5.7 years). Ultimately, this depends on operating conditions, 
and actual results may vary. If the fan runs less than approximately 4500 RPM at full power, then there may be a 
blockage or the fan may need to be replaced. The fan is secured by four nylon screws and can be easily removed. 
When a replacement fan is installed, verify that it is oriented the correct way, with air being pulled upwards 
through the shield towards the hat. Ensure that the replacement fan is wired correctly as outlined in the 
OPERATION AND MEASUREMENT section. 

Inspect the shield regularly and clean as needed to maintain optimum performance. If the shield or hat is dirty, 
wipe the surface with a moist rag to remove buildup and return the surface to its original condition. Check all 
mounting nuts and tighten if they are loose. 

For optimal accuracy, check the yellow-bead thermistor every three months for dust/dirt buildup. If needed, twist 
and pull out the sensor port adapter, then wipe the thermistor tip clean. Be sure to return the thermistor to its 
proper position as outlined in the DEPLOYMENT AND INSTALLATION section. 

Apogee ST-110 temperature sensors are factory verified to ensure accuracy. Sensors are compared for absolute 
temperature against the mean of two reference PRTs in a constant temperature bath, over a range of 
approximately -35 to 60 C.  

EE08 probes are rugged and weatherproof and designed for air temperature and relative humidity measurements 
inside radiation shields. When probes are not in use, it is recommended they be removed from the measurement 
environment, cleaned, and stored. EE08 probes, especially the filter cap, should be periodically cleaned to remove 
all dust and debris. Additional filter caps are available from Apogee to replace clogged filter caps. For more 
information about filter cap replacement, call or email Apogee technical support: 435.245.8012, 

[email protected]

. 

EE08 probes are factory calibrated and preprogrammed and come with a generic calibration factor (see Sensor 
Calibration in OPERATION AND MEASUREMENT section). A custom calibration can be derived by comparing the air 
temperature or relative humidity measurements from the probe to reference air temperature or relative humidity 
measurements. Probes can also be recalibrated at the factory. Recalibration is recommended every two years. 
Recalibration information is found on the Apogee webpage

https://www.apogeeinstruments.com/recalibration-

and-repairs/

Questions about recalibration can be emailed to

[email protected]

. 

 

Summary of Contents for TS-100

Page 1: ...21 WEST 1800 NORTH LOGAN UTAH 84321 USA TEL 435 792 4700 FAX 435 787 8268 WEB APOGEEINSTRUMENTS COM Copyright 2020 Apogee Instruments Inc OWNER S MANUAL FAN ASPIRATED RADIATION SHIELD Models TS 100 110 120 130 Rev 28 Oct 2020 ...

Page 2: ...Certificate of Compliance 3 Introduction 4 Sensor Models 5 Specifications 6 Deployment and Installation 9 Cable Connectors 10 Operation and Measurement 11 Maintenance and Recalibration 14 Troubleshooting and Customer Support 15 Return and Warranty Policy 17 ...

Page 3: ...tion for the assessment of electrical and electronic products with respect to the restriction of hazardous substances Please be advised that based on the information available to us from our raw material suppliers the products manufactured by us do not contain as intentional additives any of the restricted materials including lead see note below mercury cadmium hexavalent chromium polybrominated b...

Page 4: ...ionally accommodate PRTs with a custom adapter Apogee strongly recommends using the ST 110 and recommends the TS 110 fan aspirated radiation shield package which comes with a precision air temperature sensor model ST 110 thermistor ST 110 air temperature sensors are weatherproof have excellent long term stability and are designed for continuous air temperature measurement when housed in the TS 100...

Page 5: ...perature humidity probe if desired If the Rotronic HC2 S3 probe is used with the E3 Active Set module it must be water proofed before use with the Apogee TS 100 radiation shield The E3 Active Set module is not water proof water can infiltrate at the point where the cable enters the housing A small length of water proof heat shrink ATUM heat shrink can be used to seal the gasket where the cable ent...

Page 6: ...ere made with ST 110 thermistors in all shields In spite of careful placement of shields away from each other and the tower see photo below and filtering for wind direction these experiments indicate that there is more variability among replicate shields 0 1 C of the same model than between models of shields 0 05 C thus there is no statistical difference among the three models of shields Field tes...

Page 7: ...0 to 70 C Dimensions 220 mm height 270 mm diameter Mass 840 g Cables 5 m of shielded twisted pair wire for fan and air temperature sensors TPR jacket high water resistance high UV stability flexibility in cold conditions pigtail lead wires stainless steel 316 M8 connector located 25 cm from sensor head ST 110 Measurement Range 50 to 70 C Measurement Uncertainty 0 1 C from 0 to 70 C 0 15 C from 35 ...

Page 8: ... twisted pair wire 10 m and 20 m cables also available white TPR jacket high water resistance high UV stability flexibility in cold conditions pigtail lead wires Warranty 1 year against defects in materials and workmanship Temperature Measurement Sensor PT1000 Class A Measurement Range 40 to 80 C Output Signal range 0 to 2 5 V DC Accuracy at 20 C 0 2 C Long term Stability Less than 0 1 C per year ...

Page 9: ...adapter an example is shown on the next page If not already installed insert the adapter into the sensor port shown on next page and rotate the adapter until it matches the shield s outer contour and will not come out when pulled Look through the opening at the bottom of the shield to verify that the thermistor tip is located concentrically in the middle of the airstream If it is not remove the ad...

Page 10: ... datalogger end of the cable If you ever need a replacement cable please contact us directly to ensure ordering the proper pigtail configuration Alignment When reconnecting your sensor arrows on the connector jacket and an aligning notch ensure proper orientation Disconnection for extended periods When disconnecting the sensor for an extended period of time from a station protect the remaining hal...

Page 11: ... must have a pulse width modulation PWM output of approximately 20 kHz and a duty cycle of 50 to 100 The tachometer output is measured by a pulse counter The tachometer also requires an input voltage where maximum voltage output to the pulse counter is the input voltage to the counter The tachometer output allows monitoring of the fan to ensure it is functioning properly The following instructions...

Page 12: ...ppropriate data logger terminal following the illustrations below Thermistor Measurement devices e g datalogger controller do not measure resistance directly but determine resistance from a half bridge measurement where an excitation voltage is input across the thermistor and an output voltage is measured across the bridge resistor Wiring for EE08 SS For more information please see the EE08 SS man...

Page 13: ...er consumption can be decreased using a pulse width modulation PWM signal The PWM signal should have a frequency of approximately 20 kHz and a duty cycle of 50 to 100 where a duty cycle of 100 is full power and 50 is low power Running the fan in a low power mode reduces the current draw from 80 mA to approximately 25 mA The only way to completely stop the fan is to turn the power off The fan also ...

Page 14: ...ure sensors are factory verified to ensure accuracy Sensors are compared for absolute temperature against the mean of two reference PRTs in a constant temperature bath over a range of approximately 35 to 60 C EE08 probes are rugged and weatherproof and designed for air temperature and relative humidity measurements inside radiation shields When probes are not in use it is recommended they be remov...

Page 15: ... 350 V DC at 25 C 20 to 25 C is the approximate room temperature range Move the positive lead wire of the voltmeter to the white wire from the probe The voltage measurement should be 0 5 V DC at 20 1 25 V DC at 50 and 2 0 V DC at 80 Compatible Measurement Devices Dataloggers Controllers Meters Measurement of ST 110 thermistor resistance requires an input excitation voltage where 2 5 V DC is recomm...

Page 16: ...Apogee webpage at https www apogeeinstruments com content EplusE EE08 Temp RH Probe CR1 Effect of Cable Length Modifying Cable Length When the temperature sensor is connected to a measurement device with high input impedance sensor output signals are not changed by shortening the cable or splicing on additional cable in the field Tests have shown that if the input impedance of the measurements dev...

Page 17: ...r factory The warranty does not cover equipment that has been damaged due to the following conditions 1 Improper installation or abuse 2 Operation of the instrument outside of its specified operating range 3 Natural occurrences such as lightning fire etc 4 Unauthorized modification 5 Improper or unauthorized repair Please note that nominal accuracy drift is normal over time Routine recalibration o...

Page 18: ...anship Apogee Instruments will repair or replace the items free of charge If it is determined that your product is not covered under warranty you will be informed and given an estimated repair replacement cost PRODUCTS BEYOND THE WARRANTY PERIOD For issues with sensors beyond the warranty period please contact Apogee at techsupport apogeeinstruments com to discuss repair or replacement options OTH...

Reviews: