After clicking the ‘Recalibate’ button the user will be
prompted to cover the sensor. Place a dark cap over the
sensor and wait for the real-time PFD reading to settle
at zero. Click OK.
Uncover the sensor and wait for the PFD reading to
settle
before
entering the reference value. Click OK.
The multiplier and offset values will automatically
calculate and update in the appropriate field. Be sure to
click ‘Save’ to retain the new mu
ltiplier and offset.
Clicking
‘Data Logging’
will allow the user to log interval
measurements in a csv file while the software is open
and communicating with the sensor.
Click ‘Setup’ and the Setup Logging window appears.
Click the ‘Browse’ button to create or select a csv
file.
Select the desired sampling interval. Note that 1 second
is the minimum interval allowed. Click ‘Start’.