![Apogee Instruments SIL-111 Owner'S Manual Download Page 6](http://html.mh-extra.com/html/apogee-instruments/sil-111/sil-111_owners-manual_2954255006.webp)
SPECIFICATIONS
Calibration Traceability
Apogee SIL series infrared radiometers are calibrated to the temperature of a custom blackbody cone held at
multiple fixed temperatures over a range of radiometer (detector/sensor body) temperatures. The temperature of
the blackbody cone is measured with replicate precision thermistors thermally bonded to the cone surface. The
precision thermistors are calibrated for absolute temperature measurement against a platinum resistance
thermometer (PRT) in a constant temperature bath. The PRT calibration is directly traceable to the NIST.
SIL-111
Approximate Sensitivity
60 µV per C difference between target and detector temperature
Output from Thermopile
Approximately -3.3 to 3.3 mV for a temperature difference from -55 to 55 C
Output from Thermistor
0 to 2500 mV (typical, depends on input voltage)
Input Voltage Requirement
2500 mV excitation (typical, other voltages can be used)
Calibration Uncertainty (0 to 50 C),
when target and detector
temperature are within 20 C
0.5 C
Measurement Repeatability
Less than 0.05 C
Stability (Long-term Drift)
Less than 2 % change in slope per year when germanium filter is maintained in a
clean condition (see Maintenance and Recalibration section below)
Response Time
0.6 s, time for detector signal to reach 95 % following a step change
Field of View
22° half angle
Spectral Range
8 to 14 µm; atmospheric window (see Spectral Response below)
Operating Environment
-50 to 80 C; 0 to 100 % relative humidity (non-condensing)
Dimensions
23 mm diameter, 60 mm length
Mass
190 g (with 5m of lead wire)
Cable
5 m of six conductor, shielded, twisted-pair wire; TPR jacket (high water resistance,
high UV stability, flexibility in cold conditions); pigtail lead wires