TROUBLESHOOTING AND CUSTOMER SUPPORT 

Verify Functionality 

Pressing the power button should activate the LCD and provide a real-time UV measurement. Direct the sensor 
head toward the sun and verify the UV reading responds. Blocking all radiation from the sensor should force the 
UV reading to zero. 

Battery Life 

When the meter is maintained properly the coin cell battery (CR2320) should last for many months, even after 
continuous use. The low battery indicator will appear in the upper left hand corner of the LCD display when the 
battery voltage drops below 2.8 V DC. The meter will still function correctly for some time, but once the battery is 
drained the pushbuttons will no longer respond and any logged measurements will be lost. 

Pressing the power button to turn off the meter will actually put it in sleep mode, where there is still a slight 
amount of current draw. This is necessary to maintain the logged measurements in memory. Therefore, it is 
recommended to remove the battery when storing the meter for many months at a time, in order to preserve 
battery life. 

Low-Battery Error after Battery Replacement 

A master reset will usually correct this error, please see the master reset section for details and cautions. If a 
master reset does not remove the low battery indicator, please double check that the voltage of your new battery 
is above 2.8 V, this is the threshold for the indicator to turn on. 

Master Reset 

If a meter ever becomes non-responsive or experiences anomalies, such as a low battery indicator even after 
replacing the old battery, a master reset can be performed that may correct the problem. Note that a master reset 
will erase all logged measurements from memory.  

Step 1:  press the power button so that the LCD display is activated. 

 

Step 2: Slide the battery out of the holder, which will cause the LCD display to fade out.  

Step 3: After a few seconds, slide the battery back into the holder. 

The LCD d

isplay will flash all segments and then show a revision number (e.g. “R1.0”). This indicates the master 

reset was performed and the display should return to normal. 

Error Codes and Fixes 

Error codes will appear in place of the real-time reading on the LCD display and will continue to flash until the 
problem is corrected. Contact Apogee if the following fixes do not rectify the problem. 

Err 1:

 battery voltage out of range. 

Fix:

 replace CR2320 battery and perform master reset. 

Err 2:

 sensor voltage out of range. 

Fix: 

perform master reset. 

Err 3:

 not calibrated. 

Fix:

 perform master reset. 

Err 4:

 CPU voltage below minimum. 

Fix:

 replace CR2320 battery and perform master reset. 

Summary of Contents for MU-250

Page 1: ...RUMENTS INC 721 WEST 1800 NORTH LOGAN UTAH 84321 USA TEL 435 792 4700 FAX 435 787 8268 WEB APOGEEINSTRUMENTS COM Copyright 2021 Apogee Instruments Inc OWNER S MANUAL ULTRAVIOLET METER Models MU 250 Rev 1 Jun 2021 ...

Page 2: ...roduction 4 Sensor Models 5 Specifications 6 Deployment and Installation 7 Battery Installation and Replacement 8 Operation and Measurement 9 Apogee AMS Software 11 Maintenance and Recalibration 13 Troubleshooting and Customer Support 14 Return and Warranty Policy 16 ...

Page 3: ... electrical and electronic products with respect to the restriction of hazardous substances Please be advised that based on the information available to us from our raw material suppliers the products manufactured by us do not contain as intentional additives any of the restricted materials including lead see note below mercury cadmium hexavalent chromium polybrominated biphenyls PBB polybrominate...

Page 4: ...The output can also be expressed in photon flux density units of micromoles per square meter per second µmol m 2 s 1 Typical applications of UV meters include incoming UV radiation measurement in outdoor environments or in laboratory use with artificial light sources e g germicidal lamps Apogee Instruments MU 200 series UV A meters are connected by cable to an anodized aluminum housing MU 250 Sepa...

Page 5: ...ee MU series UV meters covered in this manual are self contained and come complete with handheld meter and sensor MU 250 Sensor model number and serial number are located on a label on the backside of the handheld meter ...

Page 6: ...m Drift Less than 2 per year Non linearity Less than 1 Response Time Less than 1 ms Field of View 180 Spectral Range 300 nm to 400 nm see Spectral Response below Directional Cosine Response 2 at 45 5 at 75 zenith angle Maximum Range 104 9 W per m2 Temperature Response Less than 0 1 per C Operating Environment 0 to 50 C less than 90 non condensing relative humidity up to 30 C less than 70 non conde...

Page 7: ...he sensor to a horizontal plane In addition to leveling all sensors should also be mounted such that obstructions e g weather station tripod tower or other instrumentation do not shade the sensor The AM 310 Sensor Wand accessory incorporates a mounting fixture at the end of an extendable telescopic wand up to 33 inches 84 cm The wand is not suited for wet environments however it is excellent for g...

Page 8: ...om the meter circuit board NOTE The battery cradle can be damaged by using an incorrectly sized battery If the battery cradle is damaged the circuit board will need to be replaced To avoid this costly problem use only a CR2320 battery Battery Removal Press down on the battery with a screwdriver or similar object Slide battery out If the battery is difficult to move turn the meter on its side so th...

Page 9: ...ke the appropriate selection Once the desired mode is blinking press the mode button three more times to exit the menu Logging To choose between manual or automatic logging push the mode button twice and use the up down buttons to make the appropriate selection SMPL or LOG Once the desired mode is blinking press the mode button two more times to exit the menu When in SMPL mode press the sample but...

Page 10: ...surements and Spectral Errors Apogee Instruments model SU 200 UV A sensors are calibrated to measure ultraviolet radiation from the sun between 300 and 400 nm in Watts per square meter In addition to naturally occurring UV A radiation from the sun there are many electric light sources that emit UV A radiation e g cool white fluorescent metal halide mercury arc and germicidal lamps Although the rel...

Page 11: ...ds When the ApogeeAMS software is first opened it will show a blank screen until communication with the meter is established If you click Open Port it will say connection failed To establish communication make sure the meter is plugged into your computer using the AC 100 communication cable To connect click the dropdown menu button and COM options will appear For more details on how to figure out ...

Page 12: ...als per day Click 30 Min Avg to see the meter s 99 30 minute averages To analyze the data click on File and Save As to save the data as a csv file Or you can highlight the numbers copy and paste them into a blank Excel spreadsheet Data will need to be comma delimited ...

Page 13: ...imates and locations around the world Although the Clear Sky Calculator does not specifically report an estimated value for UV radiation it can still be used to help determine the need for UV sensor recalibration by approximating the ratio of UV to total shortwave or the ratio of UV to PPFD However due to continuous changes in atmospheric conditions and their effect on UV radiation the comparison ...

Page 14: ... section for details and cautions If a master reset does not remove the low battery indicator please double check that the voltage of your new battery is above 2 8 V this is the threshold for the indicator to turn on Master Reset If a meter ever becomes non responsive or experiences anomalies such as a low battery indicator even after replacing the old battery a master reset can be performed that ...

Page 15: ...rcuit board of the meter Care should be taken to remove the back panel of the meter in order to access the board and splice on the additional cable otherwise two splices would need to be made between the meter and sensor head See Apogee webpage for further details on how to extend sensor cable length http www apogeeinstruments com how to make a weatherproof cable splice ...

Page 16: ...r factory The warranty does not cover equipment that has been damaged due to the following conditions 1 Improper installation or abuse 2 Operation of the instrument outside of its specified operating range 3 Natural occurrences such as lightning fire etc 4 Unauthorized modification 5 Improper or unauthorized repair Please note that nominal accuracy drift is normal over time Routine recalibration o...

Page 17: ...anship Apogee Instruments will repair or replace the items free of charge If it is determined that your product is not covered under warranty you will be informed and given an estimated repair replacement cost PRODUCTS BEYOND THE WARRANTY PERIOD For issues with sensors beyond the warranty period please contact Apogee at techsupport apogeeinstruments com to discuss repair or replacement options OTH...

Reviews: