69
11. Specifications
Model
SA-260MA-R
Chamber Capacity (L)
24
Maximum Instrument Length (mm)
350
Maximum Load (unpouched) (g)
7,000
Maximum Load (pouched) (g)
1,300
External Dimensions (mm)
655 (D) x 533 (W) x 422 (H)
Chamber Size (mm)
260 Diameter x 450 Depth
Gross Weight (kg)
61
Voltage/Frequency/Wattage
220V-240V AC, 50/60Hz, 2670W
Fuses
15A x 2, No Fuse (circuit) Breaker
Water Reservoir Capacity (ml)
4200
Water consumption per Cycle (ml)
400 to 450
Sterilization Temperature (°C)
121 / 134
Working Environment
Indoor use;
Under 1,000m (altitude);
Temperature 5°C to 40°C;
Relative Humidity 80%RH@31°C to Relative Humidity
50%RH@40°C;
Voltage fluctuation ±10 %;
Transient over voltages category II;
Pollution degree 2
Transportation Conditions
-10°C to 70°C, 10%RH to 90%RH
Storage Conditions
-10°C to 50°C, 10%RH to 70%RH
Over Pressure Protection
Safety valve 2.55 bar (2.6 kgf/cm²)
Over Pressure Indication
Pressure protection switch with warning LED
Over Temperature Indication
Independent temperature protection device with warning LED
Water Level Indication
Water level sensor in reservoir and chamber with warning LED
Door Lock Indication
Micro switch sensor with warning LED